Displaying and Monitoring the ECG
227 490 02-C
Marquette Responder® 3000
43
Activating and Deactivating Filters
Activating muscle and AC line filters makes the
displayed ECG insensitive to AC interference and
muscle tremor. However, the filters alter the ECG
signal, and a filtered signal is not suitable for
diagnosis (displayed or printed ECG). The heart
rate is always derived from the raw ECG signal.
•
Press F5
Next Menu
.
•
Using F1
Filter
, enable or disable the muscle
filter, as appropriate.
•
The AC line filter is enabled and disabled from
the setup menu.
Note
Active filters will alter the ECG signal. Switch off
the filters to obtain a diagnostic ECG.
Enabling/Disabling the QRS Beep
The QRS beep can be enabled (high, middle, low
volume) and disabled with F3
QRSPulse Beep .