OXIMETER ASSEMBLY: Theory of Operations
8-4
APEX Telemetry Transmitter
Revision D
401566-166
The following table gives examples of the oximeter readings:
The mathematics are fixed in the pulse oximeter hardware and software.
Thus, no field calibrations are needed or are possible. There are no
adjustable parts within the pulse oximeter that affect the calibration.
The function, f, depends on the wavelengths of the sensor LEDs. These
wavelengths are fixed by specified manufacturing processes and
materials. The sensors are checked for correct operation before shipping,
so no adjustment or calibration is needed or possible.
Because the pulse oximeter does all critical computations in software and
there are no critical parts to drift; no re-calibration is needed.
Example 1
Example 2
O
2
Hb = 96
O
2
Hb = 88
COHb = 0.5
COHb = 8
MetHb = 0.6
MetHb = 2
SpO
2
= 97
SpO
2
= 98
SaO
2
= 97.07
SaO
2
= 97.78
Содержание Marquette 418793-001
Страница 1: ...APEX Telemetry Transmitter Service Manual 401566 166 Revision D ...
Страница 8: ...CONTENTS vi APEX Telemetry Transmitter Revision D 401566 166 ...
Страница 100: ...TRANSMITTER PCB Schematic Diagram PN SD801278 001A 7 30 APEX Telemetry Transmitter Revision D 401566 166 Sheet 7 of 13 ...
Страница 101: ...TRANSMITTER PCB Schematic Diagram PN SD801278 001A Revision D APEX Telemetry Transmitter 7 31 401566 166 Sheet 8 of 13 ...
Страница 102: ...TRANSMITTER PCB Schematic Diagram PN SD801278 001A 7 32 APEX Telemetry Transmitter Revision D 401566 166 Sheet 9 of 13 ...
Страница 103: ...TRANSMITTER PCB Schematic Diagram PN SD801278 001A Revision D APEX Telemetry Transmitter 7 33 401566 166 Sheet 10 of 13 ...
Страница 104: ...TRANSMITTER PCB Schematic Diagram PN SD801278 001A 7 34 APEX Telemetry Transmitter Revision D 401566 166 Sheet 11 of 13 ...
Страница 105: ...TRANSMITTER PCB Schematic Diagram PN SD801278 001A Revision D APEX Telemetry Transmitter 7 35 401566 166 Sheet 12 of 13 ...
Страница 106: ...TRANSMITTER PCB Schematic Diagram PN SD801278 001A 7 36 APEX Telemetry Transmitter Revision D 401566 166 Sheet 13 of 13 ...
Страница 122: ...OXIMETER ASSEMBLY Parts Location Diagram 421191 001A 8 14 APEX Telemetry Transmitter Revision D 401566 166 Sheet 2 of 3 ...
Страница 123: ...OXIMETER ASSEMBLY Parts Location Diagram 421191 001A Revision D APEX Telemetry Transmitter 8 15 401566 166 Sheet 3 of 3 ...