Maintenance
Item
Description
7
patient cable
8
base plastic assembly
9
middle plastic assembly
10
KISS pump (option)
11
mainboard ETE assembly
12
power supply assembly
13
printer assembly
14
acquisition module assembly
Preparing System for FRU Replacement
Prior to performing any disassembly procedures, perform the following steps:
NOTE:
Take strict precautions against electrostatic discharge damage while replacing field
replaceable units.
1.
Power off the system.
2.
Disconnect the device from the AC wall outlet.
3.
Disconnect the power cord from the rear panel connector.
4.
Disconnect the patient cable from the device as described in
Patient Cable” on page 61
5.
Remove the paper tray assembly as described in
6.
Remove the battery as described in
“Replacing the Battery” on page 64
.
Replacing the Patient Cable
1.
Disconnect the system from AC power.
2.
Disconnect the patient cable from the side panel connector as shown in the
following illustration.
2028451-183E
MAC™ 1600
61
Содержание MAC 1600
Страница 18: ...Introduction 18 MAC 1600 2028451 183E...
Страница 26: ...Equipment Overview Block Diagram 26 MAC 1600 2028451 183E...
Страница 28: ...Equipment Overview 28 MAC 1600 2028451 183E...
Страница 108: ...Maintenance 108 MAC 1600 2028451 183E...
Страница 110: ...Parts Lists Field Replaceable Units FRUs MAC 1600 Upper Level Assembly PN 2032093 110 MAC 1600 2028451 183E...
Страница 111: ...Parts Lists 2028451 183E MAC 1600 111...
Страница 112: ...Parts Lists 112 MAC 1600 2028451 183E...
Страница 113: ...Parts Lists 2028451 183E MAC 1600 113...
Страница 128: ...Parts Lists 128 MAC 1600 2028451 183E...
Страница 140: ...Electromagnetic Compatibility 140 MAC 1600 2028451 183E...
Страница 145: ......