Maintenance
Electrical Safety Checks (cont'd.)
Step
Polarity
Condition
1
UUT
2
- ON
Result
Leakage
Current
Limits
1
Forward
Polarity
NC
______ µA
Pass/Fail
100 µA
2
Neutral open,
Forward
Polarity
SFC
______ µA
Pass/Fail
300 µA
3
Ground open,
Forward
Polarity
SFC
______ µA
Pass/Fail
300 µA
4
Ground open,
Reverse
Polarity
SFC
______ µA
Pass/Fail
300 µA
5
Neutral open,
Reverse
Polarity
SFC
______ µA
Pass/Fail
300 µA
6
Reverse
Polarity
NC
______ µA
Pass/Fail
100 µA
Patient Leakage Current to Ground
1
Forward
Polarity
NC
______ µA
Pass/Fail
10 µA
2
Neutral open,
Forward
Polarity
SFC
______ µA
Pass/Fail
50 µA
3
Ground open,
Forward
Polarity
SFC
______ µA
Pass/Fail
50 µA
4
Ground open,
Reverse
Polarity
SFC
______ µA
Pass/Fail
50 µA
5
Neutral open,
Reverse
Polarity
SFC
______ µA
Pass/Fail
50 µA
6
Reverse
Polarity
NC
______ µA
Pass/Fail
10 µA
Ground Continuity
Resistance
2028451-183E
MAC™ 1600
103
Содержание MAC 1600
Страница 18: ...Introduction 18 MAC 1600 2028451 183E...
Страница 26: ...Equipment Overview Block Diagram 26 MAC 1600 2028451 183E...
Страница 28: ...Equipment Overview 28 MAC 1600 2028451 183E...
Страница 108: ...Maintenance 108 MAC 1600 2028451 183E...
Страница 110: ...Parts Lists Field Replaceable Units FRUs MAC 1600 Upper Level Assembly PN 2032093 110 MAC 1600 2028451 183E...
Страница 111: ...Parts Lists 2028451 183E MAC 1600 111...
Страница 112: ...Parts Lists 112 MAC 1600 2028451 183E...
Страница 113: ...Parts Lists 2028451 183E MAC 1600 113...
Страница 128: ...Parts Lists 128 MAC 1600 2028451 183E...
Страница 140: ...Electromagnetic Compatibility 140 MAC 1600 2028451 183E...
Страница 145: ......