Maintenance
Electrical Safety Checks (cont'd.)
Step
Polarity
Condition
1
UUT
2
- ON
Result
Leakage
Current
Limits
1
AC power cord
ground prong
to exposed
metal surface
(ground lug)
N/A
______ Ω
Pass/Fail
Less than
200mΩ
1
NC = Normal Condition; SFC = Single Fault Condition; N/A = Not Applicable
2
UUT = Unit Under Test
Updating Software
Software updates are provided on an SD card. Perform a software update as
described in this section.
1.
Power up the system with the system connected to AC power.
2.
Insert the SD card with the software update in the SD card slot in the rear panel
as shown in the following illustration.
The gold contacts are face up.
3.
From the
Main Menu
, press
F5
(
System Configuration
).
104
MAC™ 1600
2028451-183E
Содержание MAC 1600
Страница 18: ...Introduction 18 MAC 1600 2028451 183E...
Страница 26: ...Equipment Overview Block Diagram 26 MAC 1600 2028451 183E...
Страница 28: ...Equipment Overview 28 MAC 1600 2028451 183E...
Страница 108: ...Maintenance 108 MAC 1600 2028451 183E...
Страница 110: ...Parts Lists Field Replaceable Units FRUs MAC 1600 Upper Level Assembly PN 2032093 110 MAC 1600 2028451 183E...
Страница 111: ...Parts Lists 2028451 183E MAC 1600 111...
Страница 112: ...Parts Lists 112 MAC 1600 2028451 183E...
Страница 113: ...Parts Lists 2028451 183E MAC 1600 113...
Страница 128: ...Parts Lists 128 MAC 1600 2028451 183E...
Страница 140: ...Electromagnetic Compatibility 140 MAC 1600 2028451 183E...
Страница 145: ......