Troubleshooting
Testing the Patient Lead Wires
Use the following procedure to test the patient leadwires:
1.
Open the
Diagnostic Tests
window as described in
Diagnostics Function” on page 33
.
2.
Connect a patient cable with lead wires to the device’s patient cable connector.
3.
Connect all leads to a patient simulator or shorting bar.
4.
Select
Patient Lead Wire Check
.
The following window opens:
5.
Press
Enter
to select
Start Test
.
The test results are displayed for each lead wire.
• If the following message is displayed, the lead wire passed the test:
Connected
.
• If the following message is displayed, the lead wire failed the test:
Disconnected
.
6.
when the test is done, press
Esc
or
F6
(
Cancel
).
7.
Replace every lead wire that failed the test.
8.
Repeat the test.
If the lead wire still fails the test, replace the acquisition board as described in
“Replacing the Acquisition Board Assembly” on page 75
Equipment Problems: ECG Data Noise
If the acquired ECG data displays unacceptable noise levels:
• Be sure the problem is not caused by poor skin preparation, placement, or condition
of the electrodes when troubleshooting noise or signal quality.
• Careful skin preparation is the key to an interference-free ECG.
• Signal quality is indicated using Hookup Advisor.
46
MAC™ 1600
2028451-183E
Содержание MAC 1600
Страница 18: ...Introduction 18 MAC 1600 2028451 183E...
Страница 26: ...Equipment Overview Block Diagram 26 MAC 1600 2028451 183E...
Страница 28: ...Equipment Overview 28 MAC 1600 2028451 183E...
Страница 108: ...Maintenance 108 MAC 1600 2028451 183E...
Страница 110: ...Parts Lists Field Replaceable Units FRUs MAC 1600 Upper Level Assembly PN 2032093 110 MAC 1600 2028451 183E...
Страница 111: ...Parts Lists 2028451 183E MAC 1600 111...
Страница 112: ...Parts Lists 112 MAC 1600 2028451 183E...
Страница 113: ...Parts Lists 2028451 183E MAC 1600 113...
Страница 128: ...Parts Lists 128 MAC 1600 2028451 183E...
Страница 140: ...Electromagnetic Compatibility 140 MAC 1600 2028451 183E...
Страница 145: ......