Location
Constraint Name
Parameter
Specification
Comment
Exam room
minimum rail length past
iso-centre towards foot end
Length
1.3 m (4 ft, 7 in)
LP4 parking at patient foot end not
allowed
For exam and control rooms, several configurations are possible.
The pre–installation instructions shall define a room layout where the location of the remote
controls versus the moving parts of equipment. This layout shall define the maximum distance
between the remote control location and equipment and the axis of the equipment (L axis) versus
the remote controls (dead angles concern).
WARNING
CARRIAGE COVERS CAN ENCLOSE DUST PARTICLES. CARE SHOULD BE
TAKEN TO AVOID PROPAGATION.
IT IS RECOMMENDED TO AVOID DIRECT AIR FLOW BETWEEN LP RAILS.
NOTE:
Motion controls installed in remote location from the table shall be installed at a location
where all the positioner axis are visible by the operator.
to see possible exam / control rooms layouts.
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
Direction 5177951-4-1EN, Revision 1
90
1 Room Layout Considerations