12 Emergency
12.1 Partial UPS Fluoro (option)
A Fluoro UPS (20 kVA) has been designed for Innova 2121-IQ, 3131-IQ. This partial UPS lets the
customer complete an exam in fluoro mode in case of a power failure. The autonomy provided by
this UPS is 5 minutes of fluoro every 24 hours.
12.2 Full UPS Record (compatability)
An UPS sized for a minimum of 150 kVA is required to supply the system in record mode. Such
an UPS would provide to the customer about 10 minutes of autonomy.
The UPS output shall follow the same output specifications as shown in
Optional Isolation
transformer specifications
in
.
Contact your local GE Healthcare representative for more information and/or public power
distribution analysis.
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
Direction 5177951-4-1EN, Revision 1