6 Insite
The preferred Insite connection uses a broadband modem. This connection requires a dedicated
Ethernet Jack (RJ45) that must be located less than 1 meter (3 feet) from the C1 frontal cabinet
For complete descriptions of the GEHC connectivity solutions, please refer to the Broadband
Solutions catalogue available through your local GEHC sales and service representative.
Connectivity Process and pre-installations checklists are available in the Broadband Connectivity
PIM available through your local GEHC sales and service representative.
The analog modem is also fitted inside the C1 cabinet. A dedicated phone line with a local socket
used only for a connection to a modem will preferably be located close to the cabinet. The phone
outlet must be located less than 1 meter (3 feet) from the C1 cabinet
A modem compliant to each country is supplied with the Innova System.
InSite requires an Internet Address connecting it to the Innova System. This address must be
available before installing the system. A request form has been defined. For more information,
please refer to
Chapter 8, IP Addressing Process
or contact your GE Healthcare OLC
representative.
NOTICE
The C1 frontal cabinet comes equipped with a Firewall unit. The hospital network
must be capable of connecting to this firewall. In the case that it cannot be, please
contact GE Healthcare to discuss alternatives.
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
Direction 5177951-4-1EN, Revision 1
84
6 Insite