2.3.4 Limitations Management
Adhering to the distance separation recommended in
will reduce disturbances recorded at the image level, but may not eliminate all disturbances.
However, when installed and operated as specified herein, the system will maintain its basic safety
and essential performance by continuing to acquire, display, and store diagnostic quality images
safely.
For example, a 1 W mobile phone (800 MHz to 2.5 GHz carrier frequency) shall be put 23 meters
) apart from the Innova system (in order to avoid images interferences risks).
2.3.5 Use Limitation
The use of accessories, transducers, and cables other than those specified may result in degraded
ELECTROMAGNETIC COMPATIBILITY of the Innova system
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
Direction 5177951-4-1EN, Revision 1
Chapter 3 Physical Requirements of room
63
Содержание Innova 2121-IQ Biplane
Страница 251: ......