Table 3-12: D distance of
ANGIO / CARDIO
CARDIO / NEURO
Omega IV Compact
NA
1395 mm (54.9 in)
Omega V Long
1278 mm (50.3 in)
1395 mm (54.9 in)
Omega V non motorized Long
1278 mm (50.3 in)
1395 mm (54.9 in)
NOTICE
Due to the plastic bushing used in the USA to protect cables from the sharp edges
of conduits it is necessary to place the cable conduit inside the table cable access
opening but the height of the outcoming conduit plus bushing is limited to 12 mm
(1/2 in).
NOTE:
Refer to table
Chemical anchors Pull out efforts and recommendations
in
for pull out effort on each fixation bolts.
3.3.5 Water Pipe Requirements
Illustration 3-7: Water Conduit location
NOTICE
In some countries, depending on local regulations, it may be forbidden to run
electrical cables and water pipes in the same conduit. In this case, two separate
conduits are required.
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
Direction 5177951-4-1EN, Revision 1
Chapter 3 Physical Requirements of room
71