18 Physical Runs
NOTICE
All lengths of cable are:
•
in useable meter when you look at group level, or
•
in meters (connector to connector) when you look at the cable level.
Suspension are always pre–cabled with 36 meter cables of connector to connector Monitor options
includes their 24 meters length for cabling
In case 24 m required, they have to be added in the order and replaced on site.
For a description of how to use the following 5 cable group schematics, see below:
Illustration 6-19: Description of cable group diagrams
Cable length data is as follows:
•
Cable Length = the total cable length, connector to connector (example above is 24 meters).
•
X + Y = used length for connection within system (example above is 5 meters).
•
Cable Length - (X + Y) = available length for conduit run (example above is 19 meters).
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
Direction 5177951-4-1EN, Revision 1
180
18 Physical Runs