1.5 Innova Imaging System
An Innova LC/LP System is managed and controlled by a System including:
•
Two C1 cabinets, inside which are located the RTAC, the DL, (DL in frontal C1 cabinet only),
the modem, the HUB, the KVM,
•
C2 LC/LP positioners control cabinet,
•
VCIM console,
•
Revolution Digital Detector mounted on Innova LC positioner,
•
Revolution Digital Detector mounted on Innova LP4 positioner,
•
Two external Digital Detector conditioners are mandatory.
•
A 3kVA cabinet UPS to be installed close to PDB (Main Disconnect Panel) whether the Fluoro
UPS option is installed or not.
•
LCD NEC 19” 1980 SXi Color Monitor.
•
LCD 19” EIZO GmbH SMD 19100G Color Monitor.
•
LCD 18” EIZO GmbH (formerly SIEMENS) SMD 18101G B&W Monitor.
•
LCD 19” EIZO GmbH SMD 19100G B&W Monitor.
1.6 Monitor suspensions
The common type of this suspension is an XT inboard monitor bridge.
A monitor frame support receiving 6 monitors: 4 B&W Monitor and 2 Color Monitor.
A monitor frame support receiving 8 monitors: 4 B&W Monitor and 4 Color Monitor or 6 B&W
Monitor and 2 Color Monitor.
In the Control Area, a Gyro TV® wall mount can support the optional 48 cm (19") LCD monitors.
1.7 DL and associated devices
An Innova System uses the DL Digital Imaging system.
•
The C1 Frontal Cabinet contains:
○
DL Computer,
○
Modem,
○
Firewall unit,
○
Ethernet switch,
○
KVM–local,
○
Video Splitter.
•
The user area is made of:
○
DL color LCD monitor (19”),
○
Keyboard,
○
Mouse,
○
Keypads for user dialogue,
○
Repeater monitors.
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
Direction 5177951-4-1EN, Revision 1
Chapter 2 Innova System
31
Содержание Innova 2121-IQ Biplane