Zone 2: Variable lighting without specular reflection on display and monitor screens. Indirect
lighting but preferable from ceiling. Automatically cut by the Innova system.
Zone 3: Variable lighting without artifact on display and monitor screens. manually cut.
Zone 4: Normal lighting manually cut. (For maintenance).
NOTICE
Zone 5: Nothing is permitted in this zone, that means:
- no mounting hardware can protrude below the finished ceiling height (top surface
of LP4 stationary rails), such as Unistrut mounting bolts, support brackets,
sprinklers, air vents, etc.
1.7 Preference Cabinet locations
NOTE:
This is applicable for all types of cabinets (including PDB).
The clear width of the service area in front of the cabinet doors to insure electrical safety shall be
at least 0.9 m. In cases where 2 cabinets are installed face to face (both sides of the access way),
the clear width shall be at least 1.2 m.
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
Direction 5177951-4-1EN, Revision 1
Chapter 4 Room Layouts