8-78
Section 8-6 - Mechanical Parts- Replacement Procedures
D
IRECTION
5771498-100, R
EVISION
6
V
ENUE
™ S
ERVICE
M
ANUAL
P R E L I M I N A R Y
Section 8-6
Mechanical Parts- Replacement Procedures
8-6-1
MPB Front Metal Door Replacement Procedure
8-6-1-1
Tools
Appropriate Phillips screwdriver.
8-6-1-2
Time Required
5 min
8-6-1-3
Preparations
1.) Shut down the Venue™ ultrasound unit, as described in
.
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-6-1-4
MPB Front Metal Door Removal Procedure
1.) Remove all accessories.
2.) Remove the Lower Front eTower cover.
3.) Release four captive screws and remove the MPB Front Metal Door.
FRU Part #
Refer to
•
Accessories - Replacement Procedures
•
Lower Front eTower Cover Removal Procedure
Figure 8-63 Removing MPB Front Metal Door