8-182
Section 8-8 - Peripherals Replacement/Installation Procedures
D
IRECTION
5771498-100, R
EVISION
6
V
ENUE
™ S
ERVICE
M
ANUAL
P R E L I M I N A R Y
8-8-5
ECG Replacement Procedure
8-8-5-1
Tools
Appropriate Phillips screwdriver
8-8-5-2
Time Required
15 min
8-8-5-3
Preparations
1.) Shut down the Venue™ ultrasound unit, as described in
.
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-8-5-4
ECG Removal Procedure
1.) Remove the following covers: Lower Front eTower, and Upper Front eTower.
2.) Release the screw that secures the ECG module in place.
FRU Part #
Refer to
.
•
Lower Front eTower Cover Removal Procedure
•
Upper eTower Front Cover Removal Procedure
Figure 8-204 Releasing ECG Securing Screw
Содержание H45051VN
Страница 477: ......