8-86
Section 8-6 - Mechanical Parts- Replacement Procedures
D
IRECTION
5771498-100, R
EVISION
6
V
ENUE
™ S
ERVICE
M
ANUAL
P R E L I M I N A R Y
8-6-4
Articulated Arm Replacement Procedure
8-6-4-1
Tools
Allen key 2.5 mm.
Snap ring (circlip) pliers
8-6-4-2
Time Required
10 min
8-6-4-3
Preparations
1.) Shut down the Venue™ ultrasound unit, as described in
.
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-6-4-4
Arm Removal Procedure
1.) Remove all accessories.
2.) Remove the following covers: RS Probe Cover, Lower Front eTower cover, Left Side eTower cover,
Right Side eTower cover and Riser Thermal Cover.
3.) Remove cockpit (monitor).
4.) Carefully raise the arm to its up position.
FRU Part #
Refer to
•
Accessories - Replacement Procedures
•
RS Probe Cover Removal Procedure
•
Lower Front eTower Cover Removal Procedure
•
Right Side eTower Cover Removal Procedure
•
Left Side eTower Cover Removal Procedure
•
Riser Thermal Cover Removal Procedure
•