2-2-5-5
Power Stability Requirements
•
Voltage drop-out
Max 10 msec
•
Power Transients
The Venue™ is fully compliant with the following standard:
EN/IEC 60601-1-2 sections 36.202.1j, 7a(2).
2-2-6
EMI Limitations
Ultrasound systems are susceptible to Electromagnetic Interference (EMI) from radio frequencies,
magnetic fields, and transients in the air or wiring. They also generate EMI. The Venue™ system
complies with limits as stated on the EMC label. However there is no guarantee that interference will
not occur in a particular installation.
NOTE:
Possible EMI sources should be identified before the unit is installed, and should not be on the same
line as the ultrasound system. A dedicated line should be used for the ultrasound system.
Electrical and electronic equipment may produce EMI unintentionally as the result of a defect. Some of
these sources include:
•
medical lasers
•
scanners
•
cauterizing guns
•
computers
•
monitors
•
fans
•
gel warmers
•
microwave ovens
•
light dimmers
•
mobile phones
•
in-house wireless phones (DECT phones)
•
wireless computer keyboard and mouse
•
air conditioning system
•
High Frequency (HF) surgery equipment
•
general AC/DC adapters
The presence of a broadcast station or broadcast van may also cause interference.