Battery - General Safety Guidelines
The lithium ion rechargeable battery packs provides power to the Venue™ system whenever an AC
power source is not available.
Venue™ ultrasound scanner is supplied with two battery packs installed in the battery bay as standard.
Each battery pack includes three lithium ion batteries bounded in plastic enclosure.
The Venue™ has built-in charger functionality and switches automatically from battery operation to AC
operation and
vice versa
.
When shutting down the system, leave the main power cable connected to keep the battery fully charged.
NOTE:
Before removing or inserting the Battery, perform system shut-down and disconnect the AC power
cable from the Venue™ ultrasound scanner.
NOTE:
The lithium ion technology used in the system's battery is significantly less hazardous to the
environment than the lithium metal technology used in some other batteries.
NOTE:
The battery is designed to be replaced every 2 years.
-
Do
not
disassemble or alter it. Charge the batteries only when the ambient temperature is
between 0 °C and 40 °C (32 °F and 104 °F) and discharge the batteries between -20 °C and
50 °C (-4 °F and 122 °F).
-
Do
not
short-circuit the battery by directly connecting the battery terminals with metal objects.
-
Do
not
heat the battery or incinerate.
-
Do
not
expose the battery to temperature over 60 °C (140 °F). Keep it away from fire and
other heat sources.
-
Do
not
charge the battery near a heat source, e.g. fire or heaters.
-
Do
not
leave the battery in direct sunlight.
-
Do
not
pierce the battery with a sharp object, hit it, or step on it.
-
Do
not
use a damaged battery. Do not solder a battery.
-
Do
not
connect the battery to an electrical outlet.
-
Do
not
immerse the battery in water or allow it to get wet.
-
Do
not
put the battery into a microwave oven or pressurized container. If the battery leaks or
emits an odor, remove it from all possible flammable sources.
-
If the battery emits an odor or heat, is deformed or discolored, or in a way appears abnormal
during use, recharging or storage, immediately remove it and stop using it.
-
If you have any questions about the battery, consult GE or your local representative.
Recommended storage conditions of battery pack:
Short term
(less than one month):
0 °C (32 °F)
to 50 °C (122 °F)
Long term
(more than three months): -20 °C (-4 °F) to 20 °C (68 °F).
CAUTION
THE BATTERY IS DESIGNED TO WORK WITH VENUE™ SYSTEMS ONLY.
ONLY USE THE BATTERIES AUTHORIZED BY GE.
Содержание H45051VN