Chapter 5 - Venue™Components and Function (Theory)
5-23
P R E L I M I N A R Y
For information on cables connected to the BEP Module, see
BEP-to-Front End (T-CFE) Interface
is via the
Data and Control Cable
that contains the following:
•
PCI-e channels
•
Control and Status signals between the BEP and the Front End.
BEP Interface Panel
- this panel allows external connections of multiple interfaces by extension cables
that connects the internal ports to the IP panel.
•
The BEP Interface Panel is located on the rear side of the Venue™ ultrasound scanner, and
contains the following connectors:
•
HDMI port - supports connection of external monitor
•
2 USB 3.0 ports - support for external USB devices such as Bar-code reader etc.
•
RJ45 - Ethernet port: supports wired network connection.
Figure 5-17 BEP Board Interfaces