2-2
Section 2-2 - Console Requirements
D
IRECTION
5771498-100, R
EVISION
6
V
ENUE
™ S
ERVICE
M
ANUAL
P R E L I M I N A R Y
Section 2-2
Console Requirements
2-2-1
Unit Environmental Requirements
NOTE:
The Venue™ system may be operated at an altitude of up to 3000 meters (9842 ft).
NOTE:
After a long period of storage, or after transportation of the system with the monitor in the folded-down
position (transportation mode), it is highly recommended to place the monitor in the upright position -
and to leave it in this position for a period of longer than 1 hour before use. This will enable it to properly
adjust to the environmental conditions.
2-2-2
Cooling Requirements
The cooling requirement for the Venue™ ultrasound unit environment is 2000 BTU/hr. This figure does
not include the cooling required for lights, people, or other equipment in the room.
NOTE:
Each person in the room places an additional 300 BTU/hr demand on the environmental cooling.
2-2-3
Lighting Requirements
Bright light is needed for Ultrasound system installation, updates and repairs. However, operator and
patient comfort may be optimized if the room light is subdued and indirect. Therefore a combination
lighting system (dim/bright) is recommended. Keep in mind that lighting controls and dimmers can be a
source of EMI which could degrade image quality. These controls should be selected to minimize
possible interference.
2-2-4
Time and Manpower Requirements.
Table 2-2
Environmental Requirements
Requirement
Temperature
Relative Humidity
(non-condensing)
Air Pressure
Operational
+10 —
+40
o
C (50 — 104
o
F)
30 — 85%
700 — 1060 hPa
Storage
-20 — +60
o
C
(-4 — 140
o
F)
10 — 70%
700 — 1060 hPa
Transport
-20 — +60
o
C
(-4 — 140
o
F))
10 — 70%
700— 1060 hPa
CAUTION
IF THE SYSTEM HAS BEEN IN STORAGE OR HAS BEEN TRANSPORTED, PLEASE SEE THE
ACCLIMATION REQUIREMENTS BEFORE POWERING ON AND/OR USING THE SYSTEM.
Refer to the
CAUTION
Two people are required to unpack the Venue™ ultrasound unit; at least two
people must be available to roll the system down the wheeling ramp. Attempts to
move the system considerable distances (or on an incline) by one person alone,
could result in personal injury, and/or damage to the system.
Содержание H45051VN
Страница 477: ......