8-40
Section 8-4 - System Modules- Replacement Procedures
D
IRECTION
5771498-100, R
EVISION
6
V
ENUE
™ S
ERVICE
M
ANUAL
P R E L I M I N A R Y
8-4-6
PSU Module Replacement Procedure
8-4-6-1
Tools
Appropriate Phillips screwdriver.
8-4-6-2
Time Required
15 min
8-4-6-3
Preparations
1.) Shut down the Venue™ ultrasound unit, as described in
.
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-4-6-4
PSU Module Removal Procedure
1.) Remove all accessories.
2.) Remove the following covers: RS Probe Cover, Lower Front eTower cover, Right Side eTower
Cover and the Left Side eTower Cover:.
3.) Remove the MPB door.
4.) Remove the two battery packs
FRU Part #
Refer to
.
•
Accessories - Replacement Procedures
•
RS Probe Cover Removal Procedure
•
Lower Front eTower Cover Removal Procedure
•
Right Side eTower Cover Removal Procedure
•
Left Side eTower Cover Removal Procedure
•
MPB Door Cover Removal Procedure
•
Содержание H45051VN