Chapter 5: Troubleshooting
Troubleshooting Symptoms
Service Manual
5-7
S29
Device will not power up, alarm
sounds.
Defective scale
Disconnect scale connector at
probe panel. Power cycle the
device. If error clears replace
the scale.
Defective scale harness
Disconnect scale harness from
J7 on the control board. Power
cycle the device. If error clears
replace the scale harness.
Defective Servo Oxygen
board
Disconnect the RS485 harness
from J8 on the Servo Oxygen
board. Power cycle the device.
If the error clears, replace the
Servo Oxygen Board.
Defective Servo Oxygen
cable
Disconnect RS485 bus cable
from J6 on the control board.
Power cycle the device. If the
error clears, replace the RS485
cable going to the Servo
Oxygen board.
Defective PF Battery
Remove the electrical
enclosure cover and
disconnect the PF Battery.
Power down and power up
again. If the device powers up,
replace the 9V PF Battery
(66600-1024-600).
Defective power supply
Check power supply voltages.
Refer to the Tips section (
).
Defective power harness
to SBC
Verify power supply voltages at
J17 on SBC.
•
+5V 1-2
•
+12V 1-3
•
+5STBY 1-4
Defective SBC
Disconnect ribbon cable from
J6 on control board. Power up
device. Giraffe logo should
appear on display. After about
70 seconds the audible alarm
sounds. If the logo does not
appear replace the SBC.
Defective control board
Replace control board.
S30
Patient Hot or patient Cold
alarms activated while device
is in Manual Mode.
MANUAL TEMP alarm is
selected on the operator
Setup screen.
Select OFF for MANUAL TEMP
alarm on Setup screen.
ID
Symptom
Possible Cause(s)
Action(s)
Содержание Giraffe OmniBed Carestation CS1
Страница 4: ...RH 2 Service Manual ...
Страница 24: ...Service Language Disclaimer viii Service Manual ...
Страница 40: ...Safety and Regulatory Information User Responsibility xxiv Service Manual ...
Страница 44: ...Chapter 1 Functional Description Control Board 1 2 Service Manual FIGURE 1 1 Block Diagram ...
Страница 74: ...Chapter 3 Calibration and Maintenance Line Voltage Calibration 3 12 Service Manual ...
Страница 82: ...Chapter 4 Understanding the Service Interface Upgrade 4 8 Service Manual ...
Страница 110: ...Chapter 5 Troubleshooting Diagnostic Readings 5 28 Service Manual ...
Страница 211: ...Chapter 7 Repair Procedures Display Harness Connections Service Manual 7 75 ...
Страница 212: ...Chapter 7 Repair Procedures Display Harness Connections 7 76 Service Manual ...
Страница 272: ...Chapter 8 Illustrated Parts Options 8 60 Service Manual FIGURE 8 36 Silo Support Assembly 6600 0853 800 ...
Страница 290: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 78 Service Manual FIGURE 8 45 Wiring Diagram Control Panel ...
Страница 291: ...Chapter 8 Illustrated Parts Wiring Diagrams Service Manual 8 79 FIGURE 8 46 Wiring Diagram Incubator and Warmer Heater ...
Страница 292: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 80 Service Manual FIGURE 8 47 Wiring Diagram Servo Humidifier ...
Страница 293: ...Chapter 8 Illustrated Parts Wiring Diagrams Service Manual 8 81 FIGURE 8 48 Wiring Diagram Servo Control Oxygen ...
Страница 294: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 82 Service Manual ...
Страница 296: ...Appendix A Compartment and Skin Probe Characteristics A 2 Service Manual ...
Страница 304: ...Appendix C RS 232 Serial Data Nurse Call C 4 Service Manual ...