Chapter 6: Alarms and System Failure Codes
Alarms
6-2
Service Manual
6.1.3 Patient Hot - Check Probe 1
6.1.4 Patient Cold - Check Probe 1
6.1.5 Canopy Pedal Disabled
Cause or Circumstances
Failure
Action
The Patient Hot alarm activates
when the temperature
measured by the skin
temperature probe on the
patient skin is 1.0°C above the
set temperature. The alarm
limit can be changed to 0.5 °C
in the set-up menu. The Skin
Temperature alarm must be
turned on for this alarm to
trigger while in Manual Mode.
Baby Mode set temperature is
incorrect for patient.
External heat such as
phototherapy or sunlight.
Skin temperature probe is
disconnected from skin.
Respond per clinical protocol.
Check the attachment of the skin
temperature probe on the patient’s
skin.
Cause or Circumstances
Failure
Action
The Patient Cold alarm
activates when the
temperature measured by the
skin temperature probe on the
patient skin is 1.0°C below the
set temperature.
he alarm limit can be changed
to 0.5 °C in the set-up menu.
The Skin Temperature alarm
must be turned on for this
alarm to trigger while in
Manual Mode.
Baby Mode set temperature is
incorrect for patient.
Probe disconnected from skin.
The patient is cold.
Skin Temperature alarm is On.
Respond per clinical protocol.
If not in Baby Mode, switch to Baby
Mode.
Confirm Skin Temperature setting
in Setup.
Check the attachment of the skin
temperature probe on the patient’s
skin.
Cause or Circumstances
Failure
Action
The Canopy Pedal Disabled
alarm activates if the pedal has
been disabled on either the
Service screen or the Setup
screen and the pedal is
pressed.
Canopy pedal is disabled in
service mode.
Canopy pedal is disabled in
Setup screen.
Release the pedal.
If the canopy pedal is disabled in
the service screen and you wish to
use it, contact authorized service
personnel to reactivate it on the
service screen.
Содержание Giraffe OmniBed Carestation CS1
Страница 4: ...RH 2 Service Manual ...
Страница 24: ...Service Language Disclaimer viii Service Manual ...
Страница 40: ...Safety and Regulatory Information User Responsibility xxiv Service Manual ...
Страница 44: ...Chapter 1 Functional Description Control Board 1 2 Service Manual FIGURE 1 1 Block Diagram ...
Страница 74: ...Chapter 3 Calibration and Maintenance Line Voltage Calibration 3 12 Service Manual ...
Страница 82: ...Chapter 4 Understanding the Service Interface Upgrade 4 8 Service Manual ...
Страница 110: ...Chapter 5 Troubleshooting Diagnostic Readings 5 28 Service Manual ...
Страница 211: ...Chapter 7 Repair Procedures Display Harness Connections Service Manual 7 75 ...
Страница 212: ...Chapter 7 Repair Procedures Display Harness Connections 7 76 Service Manual ...
Страница 272: ...Chapter 8 Illustrated Parts Options 8 60 Service Manual FIGURE 8 36 Silo Support Assembly 6600 0853 800 ...
Страница 290: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 78 Service Manual FIGURE 8 45 Wiring Diagram Control Panel ...
Страница 291: ...Chapter 8 Illustrated Parts Wiring Diagrams Service Manual 8 79 FIGURE 8 46 Wiring Diagram Incubator and Warmer Heater ...
Страница 292: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 80 Service Manual FIGURE 8 47 Wiring Diagram Servo Humidifier ...
Страница 293: ...Chapter 8 Illustrated Parts Wiring Diagrams Service Manual 8 81 FIGURE 8 48 Wiring Diagram Servo Control Oxygen ...
Страница 294: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 82 Service Manual ...
Страница 296: ...Appendix A Compartment and Skin Probe Characteristics A 2 Service Manual ...
Страница 304: ...Appendix C RS 232 Serial Data Nurse Call C 4 Service Manual ...