Appendix B: Specifications
Service Manual
B-3
Irradiance
During the ten minute warm up period, the radiant heater is at 100% power (~27 mW/cm
2
). After the radiant heater
has been running in warm up mode for ten minutes, the heater power setting drops to 25% (less than 10 mW/cm
2
).
Note:
The 25% setting can be changed in Service Mode.
Weight Scale Performance
Functional Range
300g - 8 kg
± 10g
Accuracy
Resolution
10g (Factory setting) or 5g (user selectable on non-EU scales)
The resolution of scales distributed in the EU region vary depending on the
scale revision:
•
Scales with revision 2.72 only have a resolution option of 10g.
•
Scales with revision 2.86 have a resolution of 5g for weights up to 5 kg and
a resolution 10g for weights from 5 kg to 8 kg.
•
The resolution on EU scales with software 2.86 and above is not user
selectable.
Note:
For the European Union (EU) and the European Free Trade Association (EFTA) member states in Europe, check
with your local regulations with respect to the Non-Automated Weighing Instrument (NAWI) directive as to the
process and frequency for scale calibration. The resolution of scales distributed in this EU region only have a
resolution option of 10g.
Servo Oxygen
The time from turning on the servo control oxygen system until the system is fully operational is less than 10 seconds.
Control Range
21% - 65%
Resolution
1%
Accuracy*
5% (< 0.02% drift over 24 hours)
Rise Time
≤
10 minutes from 21% to 5% below set point (in Whisper Quiet™ mode)
Recovery from Opening Porthole
≤
5 minutes from closing porthole to 5% below set point (in Whisper Quiet™
mode)
Inlet Pressure
310 kPa (45 psi) - 620 kPa (90 psi)
Effects of Barometric Pressure: The Servo O2 system is not equipped with
automatic barometric pressure compensation.
Between Servo O2 system calibrations, a 1 kPa barometric pressure increase
will increase the oxygen concentration reading by 0.2% O
2
.
Note:
The above item noted with an asterisk (*) indicates essential performance.
Содержание Giraffe OmniBed Carestation CS1
Страница 4: ...RH 2 Service Manual ...
Страница 24: ...Service Language Disclaimer viii Service Manual ...
Страница 40: ...Safety and Regulatory Information User Responsibility xxiv Service Manual ...
Страница 44: ...Chapter 1 Functional Description Control Board 1 2 Service Manual FIGURE 1 1 Block Diagram ...
Страница 74: ...Chapter 3 Calibration and Maintenance Line Voltage Calibration 3 12 Service Manual ...
Страница 82: ...Chapter 4 Understanding the Service Interface Upgrade 4 8 Service Manual ...
Страница 110: ...Chapter 5 Troubleshooting Diagnostic Readings 5 28 Service Manual ...
Страница 211: ...Chapter 7 Repair Procedures Display Harness Connections Service Manual 7 75 ...
Страница 212: ...Chapter 7 Repair Procedures Display Harness Connections 7 76 Service Manual ...
Страница 272: ...Chapter 8 Illustrated Parts Options 8 60 Service Manual FIGURE 8 36 Silo Support Assembly 6600 0853 800 ...
Страница 290: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 78 Service Manual FIGURE 8 45 Wiring Diagram Control Panel ...
Страница 291: ...Chapter 8 Illustrated Parts Wiring Diagrams Service Manual 8 79 FIGURE 8 46 Wiring Diagram Incubator and Warmer Heater ...
Страница 292: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 80 Service Manual FIGURE 8 47 Wiring Diagram Servo Humidifier ...
Страница 293: ...Chapter 8 Illustrated Parts Wiring Diagrams Service Manual 8 81 FIGURE 8 48 Wiring Diagram Servo Control Oxygen ...
Страница 294: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 82 Service Manual ...
Страница 296: ...Appendix A Compartment and Skin Probe Characteristics A 2 Service Manual ...
Страница 304: ...Appendix C RS 232 Serial Data Nurse Call C 4 Service Manual ...