Chapter 6: Alarms and System Failure Codes
System Failure Codes
6-12
Service Manual
6.1.22 Up Pedal Pressed Alarm
6.1.23 Replace Clock Battery
When the clock battery voltage is too low to keep proper time, the following message appears:
Call service to replace the clock battery. Continued usage will not retain patient settings or date and
time.
6.2 System Failure Codes
The following sections describe the meaning of each system failure code for the device Each table lists the
possible causes and the corresponding recommended actions, in the order in which they should be tried.
6.2.1 System Failure 0
System failure 0 indicates communication error between the SBC and the control board.
Cause or Circumstances
Failure
Action
The bed up pedal switch was
pressed and held on during
power up.
None
When the bed up pedal switch is
released, the alarm deactivates. If
the alarm silence button is pressed
or the pedal remains on for 100
seconds, the Bed Up Pedal Failure
alarm activates and the Bed Up
Pedal Switches are disabled. Shut
down the device and power up to
clear failure.
Other
Bed up pedal switch has failed.
Refer to section
for
troubleshooting information.
Cause(s)
Action(s)
Missing or bad clock
battery
Replace the battery on the SBC and set proper time/date in Service Mode.
Cause(s)
Action(s)
Defective scale
Disconnect the scale connector at probe panel. Power cycle the device. If
the error clears, replace the scale.
Defective scale harness
Disconnect the scale harness from J7 on the control board. (The scale may
be plugged in.) Power cycle the device. If the error clears, replace the scale
harness.
Defective Servo Oxygen
board
Disconnect the RS485 harness from J8 on the Servo Oxygen board. Power
cycle the device. If the error clears, replace the Servo Oxygen Board.
Содержание Giraffe OmniBed Carestation CS1
Страница 4: ...RH 2 Service Manual ...
Страница 24: ...Service Language Disclaimer viii Service Manual ...
Страница 40: ...Safety and Regulatory Information User Responsibility xxiv Service Manual ...
Страница 44: ...Chapter 1 Functional Description Control Board 1 2 Service Manual FIGURE 1 1 Block Diagram ...
Страница 74: ...Chapter 3 Calibration and Maintenance Line Voltage Calibration 3 12 Service Manual ...
Страница 82: ...Chapter 4 Understanding the Service Interface Upgrade 4 8 Service Manual ...
Страница 110: ...Chapter 5 Troubleshooting Diagnostic Readings 5 28 Service Manual ...
Страница 211: ...Chapter 7 Repair Procedures Display Harness Connections Service Manual 7 75 ...
Страница 212: ...Chapter 7 Repair Procedures Display Harness Connections 7 76 Service Manual ...
Страница 272: ...Chapter 8 Illustrated Parts Options 8 60 Service Manual FIGURE 8 36 Silo Support Assembly 6600 0853 800 ...
Страница 290: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 78 Service Manual FIGURE 8 45 Wiring Diagram Control Panel ...
Страница 291: ...Chapter 8 Illustrated Parts Wiring Diagrams Service Manual 8 79 FIGURE 8 46 Wiring Diagram Incubator and Warmer Heater ...
Страница 292: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 80 Service Manual FIGURE 8 47 Wiring Diagram Servo Humidifier ...
Страница 293: ...Chapter 8 Illustrated Parts Wiring Diagrams Service Manual 8 81 FIGURE 8 48 Wiring Diagram Servo Control Oxygen ...
Страница 294: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 82 Service Manual ...
Страница 296: ...Appendix A Compartment and Skin Probe Characteristics A 2 Service Manual ...
Страница 304: ...Appendix C RS 232 Serial Data Nurse Call C 4 Service Manual ...