Chapter 6: Alarms and System Failure Codes
Alarms
Service Manual
6-5
6.1.10 Bed Raise/Lower Pedal Failure Alarm
6.1.11 Canopy Pedal Failure Alarm
Cause or Circumstances
Failure
Action
One of the bed Raise/Lower
pedal switches is shorted.
Defective bed Raise/Lower
pedal switch or harness
Check the switches on the service
screen to determine which switch is
shorted or open. To determine if a
switch is defective or the wiring has
a short take off the back cover,
disconnect the lower shroud from
the base plate, lean the shroud
forward to access the switches,
and check continuity at the switch
connectors. Check connections
before reassembling.
Defective relay board
Defective relay board
If the alarm persists, replace the
relay board.
Cause or Circumstances
Failure
Action
One of the canopy pedal
switches is shorted. (The alarm
appears about 100 seconds
after the short.)
Defective canopy pedal switch
or harness
Check the switches on the service
screen to determine which is
shorted. Refer to the Tips section
(4.6.9). To determine if a switch is
defective or the wiring has a short
take off the back cover, disconnect
the lower shroud from the base
plate, lean the shroud forward to
access the switches, and check
continuity at the switch connectors.
Check connections before
reassembling.
One of the canopy hand
switches is shorted.
Defective canopy hand switch
or harness
Replace one switch. If the error
persists, replace the other switch.
Содержание Giraffe OmniBed Carestation CS1
Страница 4: ...RH 2 Service Manual ...
Страница 24: ...Service Language Disclaimer viii Service Manual ...
Страница 40: ...Safety and Regulatory Information User Responsibility xxiv Service Manual ...
Страница 44: ...Chapter 1 Functional Description Control Board 1 2 Service Manual FIGURE 1 1 Block Diagram ...
Страница 74: ...Chapter 3 Calibration and Maintenance Line Voltage Calibration 3 12 Service Manual ...
Страница 82: ...Chapter 4 Understanding the Service Interface Upgrade 4 8 Service Manual ...
Страница 110: ...Chapter 5 Troubleshooting Diagnostic Readings 5 28 Service Manual ...
Страница 211: ...Chapter 7 Repair Procedures Display Harness Connections Service Manual 7 75 ...
Страница 212: ...Chapter 7 Repair Procedures Display Harness Connections 7 76 Service Manual ...
Страница 272: ...Chapter 8 Illustrated Parts Options 8 60 Service Manual FIGURE 8 36 Silo Support Assembly 6600 0853 800 ...
Страница 290: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 78 Service Manual FIGURE 8 45 Wiring Diagram Control Panel ...
Страница 291: ...Chapter 8 Illustrated Parts Wiring Diagrams Service Manual 8 79 FIGURE 8 46 Wiring Diagram Incubator and Warmer Heater ...
Страница 292: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 80 Service Manual FIGURE 8 47 Wiring Diagram Servo Humidifier ...
Страница 293: ...Chapter 8 Illustrated Parts Wiring Diagrams Service Manual 8 81 FIGURE 8 48 Wiring Diagram Servo Control Oxygen ...
Страница 294: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 82 Service Manual ...
Страница 296: ...Appendix A Compartment and Skin Probe Characteristics A 2 Service Manual ...
Страница 304: ...Appendix C RS 232 Serial Data Nurse Call C 4 Service Manual ...