Chapter 5: Troubleshooting
Switches/Thermostat
5-12
Service Manual
c.
Connect the negative (com) probe of multimeter to pin 1 and the positive probe to pin 4. The
voltage shall be more than 4.75V. If not, harness 6600-0701-700 is defective. Order a new harness
and replace it.
Note:
When replacing either harness, do not to pull by the cable wires, but use the connector body to
disengage or engage connectors.
Note:
I
f the voltages on pins 2, 3, 4 are all within the acceptable voltage range, then both harnesses are
good to use and the symptom is not related to harness damage.
5.3 Switches/Thermostat
Use switch status diagram on the Switches service screen to assist in troubleshooting the switches.
When closed the resistance should be less than 200 ohms.
Canopy up detect
Relay bd J36 pins 1-2
Closed when canopy is up
Canopy down detect
Relay bd J36 pins 7-8
Closed when canopy is down
Canopy middle
Relay bd J36 pins 4-5
Closed momentarily during
transition
Heater doors open
Relay bd J31 pins 2-3
Closed when both heater doors
are open
Heater doors closed
Relay bd J31 pins 1-3
Closed when both heater doors
are closed
Humidity reservoir
Relay bd J32 pins 2-3
Closed when reservoir is closed
Add water thermostat
Relay bd J32 pins 1-3
Opens when reservoir needs
water
Canopy foot control
Left or Right
Relay bd J40 pins 3-4
Closed when either switch is
pressed
Elevating Base Foot Control
Left or right up
Relay bd J40 pins 2-4
Closed when either switch is
pressed
Left or right down
Relay bd J40 pins 1-4
Closed when either switch is
pressed
Canopy hand control
Left up
Relay bd J35 pins 2-4
Closed when switch is pressed
Left down
Relay bd J35 pins 3-4
Closed when switch is pressed
Right up
Relay bd J35 pins 6-5
Closed when switch is pressed
Right down
Relay bd J35 pins 7-5
Closed when switch is pressed
Содержание Giraffe OmniBed Carestation CS1
Страница 4: ...RH 2 Service Manual ...
Страница 24: ...Service Language Disclaimer viii Service Manual ...
Страница 40: ...Safety and Regulatory Information User Responsibility xxiv Service Manual ...
Страница 44: ...Chapter 1 Functional Description Control Board 1 2 Service Manual FIGURE 1 1 Block Diagram ...
Страница 74: ...Chapter 3 Calibration and Maintenance Line Voltage Calibration 3 12 Service Manual ...
Страница 82: ...Chapter 4 Understanding the Service Interface Upgrade 4 8 Service Manual ...
Страница 110: ...Chapter 5 Troubleshooting Diagnostic Readings 5 28 Service Manual ...
Страница 211: ...Chapter 7 Repair Procedures Display Harness Connections Service Manual 7 75 ...
Страница 212: ...Chapter 7 Repair Procedures Display Harness Connections 7 76 Service Manual ...
Страница 272: ...Chapter 8 Illustrated Parts Options 8 60 Service Manual FIGURE 8 36 Silo Support Assembly 6600 0853 800 ...
Страница 290: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 78 Service Manual FIGURE 8 45 Wiring Diagram Control Panel ...
Страница 291: ...Chapter 8 Illustrated Parts Wiring Diagrams Service Manual 8 79 FIGURE 8 46 Wiring Diagram Incubator and Warmer Heater ...
Страница 292: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 80 Service Manual FIGURE 8 47 Wiring Diagram Servo Humidifier ...
Страница 293: ...Chapter 8 Illustrated Parts Wiring Diagrams Service Manual 8 81 FIGURE 8 48 Wiring Diagram Servo Control Oxygen ...
Страница 294: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 82 Service Manual ...
Страница 296: ...Appendix A Compartment and Skin Probe Characteristics A 2 Service Manual ...
Страница 304: ...Appendix C RS 232 Serial Data Nurse Call C 4 Service Manual ...