Chapter 7: Repair Procedures
Radiant Heater Assembly Repair Procedures
7-40
Service Manual
3.
To replace the door cable, use a flat screwdriver blade to pry the top tang (12 o’clock position) on the
spool assembly and pull the cable off the spool toward you. Use a 2.5 mm hex key to separate the cable
from the spool.
7.6.4 Heater Door Cable Adjustment
“Figure 7-39 Heater Door Cable Adjustment”
1.
The bottom end cap on the right side upright should be removed. Position the top nut and flat washer
so the bottom of the flat washer is in the middle of the threads. Slide the heater door opening rod
through the bracket, then install the flat washer, lock washer, and nut (8 mm wrench).
2.
Use a straight blade screwdriver in the slot at the cable rod’s end to hold the rod and use the wrench to
tighten the bottom nut.
3.
Turn on the device and run the canopy to its upper travel limit, and verify that both doors fully open.
7.6.5 Spring Tube Assembly Replacement
For an exploded view of the spring tube assembly, refer to section
.
1.
Elevate the elevating base to the upper limit of travel.
2.
Elevate the canopy to the service position, about halfway up so that the black oxide shipping screw is
centered in the horizontal slot, which is about midway up on the outside of the dovetail rail.
3.
Turn off the device and unplug it from the wall.
4.
Slide up the decorative strip in the dovetail rail slot to expose lower end cap screw. (A piece of adhesive
tape on the decorative strip can help to move it.)
5.
Remove the lower end cap.
6.
Remove the hole plug from the back of East lift rail. (Refer to
“Figure 7-22 Disconnecting Heater Door
7.
Rotate the spring tube assembly until the setscrew hole and lower securing hole are visible from the
access hole in the back of the inner rail. Insert a 2 mm hex wrench through the small hole in the back of
the inner rail through the spring tube and back into the rail to temporarily secure the spring tube
assembly to the inner rail.
8.
Unscrew the setscrew in the door cable conical cap until it is possible to separate the door cable cap
from the spring tube. Do not remove the setscrew.
9.
Secure a length of string to the cross-hole in the spring tube. The string should be long enough to allow
the spring tube to be removed from the bottom of the rail assembly. Firmly secure the other end of the
string. This string will be used to lift the new spring tube assembly into place.
FIGURE 7-40. Securing String to the Spring Tube
Содержание Giraffe OmniBed Carestation CS1
Страница 4: ...RH 2 Service Manual ...
Страница 24: ...Service Language Disclaimer viii Service Manual ...
Страница 40: ...Safety and Regulatory Information User Responsibility xxiv Service Manual ...
Страница 44: ...Chapter 1 Functional Description Control Board 1 2 Service Manual FIGURE 1 1 Block Diagram ...
Страница 74: ...Chapter 3 Calibration and Maintenance Line Voltage Calibration 3 12 Service Manual ...
Страница 82: ...Chapter 4 Understanding the Service Interface Upgrade 4 8 Service Manual ...
Страница 110: ...Chapter 5 Troubleshooting Diagnostic Readings 5 28 Service Manual ...
Страница 211: ...Chapter 7 Repair Procedures Display Harness Connections Service Manual 7 75 ...
Страница 212: ...Chapter 7 Repair Procedures Display Harness Connections 7 76 Service Manual ...
Страница 272: ...Chapter 8 Illustrated Parts Options 8 60 Service Manual FIGURE 8 36 Silo Support Assembly 6600 0853 800 ...
Страница 290: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 78 Service Manual FIGURE 8 45 Wiring Diagram Control Panel ...
Страница 291: ...Chapter 8 Illustrated Parts Wiring Diagrams Service Manual 8 79 FIGURE 8 46 Wiring Diagram Incubator and Warmer Heater ...
Страница 292: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 80 Service Manual FIGURE 8 47 Wiring Diagram Servo Humidifier ...
Страница 293: ...Chapter 8 Illustrated Parts Wiring Diagrams Service Manual 8 81 FIGURE 8 48 Wiring Diagram Servo Control Oxygen ...
Страница 294: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 82 Service Manual ...
Страница 296: ...Appendix A Compartment and Skin Probe Characteristics A 2 Service Manual ...
Страница 304: ...Appendix C RS 232 Serial Data Nurse Call C 4 Service Manual ...