Chapter 7: Repair Procedures
Right Rail Assembly Repair Procedures
7-24
Service Manual
7.4.3.2 Replacing Rollers and Tension Springs
“Figure 7-26 Rollers and Tensioning Spring”
1.
The rollers can be removed one at a time by removing the screw (4 mm hex key) at their hub and pulling
off the roller and bearing. When reinstalling, torque the screws to 45 in. lbs (reference).
2.
To replace the roller tensioning spring, first use needle nose pliers to pull the tension spring off the roll
pin at the end of the lift rail. Then take off the 2 center rollers and slide the tensioning plate out of the lift
rail. Remove the old spring.
3.
Reinstall the 2 center rollers. Use a hex key in the center access slot to lift the tensioning plate up so
that you can screw the rollers into the holes in the plate. Torque rollers screws to 45 inch lbs (reference).
4.
Using pliers, pull the tensioning spring up onto the roll pin at the end of the rail.
7.4.3.3 Separating the Lift Rail from the Belt Channel
“Figure 7-25 Right Rail Disassembly”
1.
If you can access the screw head through the access hole in the lift rail go to step 2. If you can not
access the screw, push the lift rail and the belt channel together and lift the belt block up past the roll
pin so you can see the screw head through the access hole in the lift rail.
STEP 3
FIGURE 7-26. Rollers and Tensioning Spring
Bearing
Roller mounting screw
Center rollers
Tensioning plate
Roller tensioning spring
Lift rail
Содержание Giraffe OmniBed Carestation CS1
Страница 4: ...RH 2 Service Manual ...
Страница 24: ...Service Language Disclaimer viii Service Manual ...
Страница 40: ...Safety and Regulatory Information User Responsibility xxiv Service Manual ...
Страница 44: ...Chapter 1 Functional Description Control Board 1 2 Service Manual FIGURE 1 1 Block Diagram ...
Страница 74: ...Chapter 3 Calibration and Maintenance Line Voltage Calibration 3 12 Service Manual ...
Страница 82: ...Chapter 4 Understanding the Service Interface Upgrade 4 8 Service Manual ...
Страница 110: ...Chapter 5 Troubleshooting Diagnostic Readings 5 28 Service Manual ...
Страница 211: ...Chapter 7 Repair Procedures Display Harness Connections Service Manual 7 75 ...
Страница 212: ...Chapter 7 Repair Procedures Display Harness Connections 7 76 Service Manual ...
Страница 272: ...Chapter 8 Illustrated Parts Options 8 60 Service Manual FIGURE 8 36 Silo Support Assembly 6600 0853 800 ...
Страница 290: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 78 Service Manual FIGURE 8 45 Wiring Diagram Control Panel ...
Страница 291: ...Chapter 8 Illustrated Parts Wiring Diagrams Service Manual 8 79 FIGURE 8 46 Wiring Diagram Incubator and Warmer Heater ...
Страница 292: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 80 Service Manual FIGURE 8 47 Wiring Diagram Servo Humidifier ...
Страница 293: ...Chapter 8 Illustrated Parts Wiring Diagrams Service Manual 8 81 FIGURE 8 48 Wiring Diagram Servo Control Oxygen ...
Страница 294: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 82 Service Manual ...
Страница 296: ...Appendix A Compartment and Skin Probe Characteristics A 2 Service Manual ...
Страница 304: ...Appendix C RS 232 Serial Data Nurse Call C 4 Service Manual ...