Safety and Regulatory Information
Warnings
x
Service Manual
!
Warning:
Use the device only with the power cord supplied.
!
Warning:
The use of accessories, replacement parts, or power cords other than
those specified by the manufacturer may affect the performance of the device and
could result in damage to the device or unsafe operating conditions.
!
Warning:
To avoid the risk of electrical shock, this equipment must directly connect
to a supply mains with a grounded outlet or a hospital grade outlet.
!
Warning:
Cover patients’ eyes if an observation light is used for an extended time.
Prolonged exposure to an observation light can damage patient’s eyes.
!
Warning:
I
nsulation on electrical wiring can deteriorate with age. Check for brittle
or deteriorated insulation on power cord and all other electrical wiring.
!
Warning:
Do not clean or disinfect while the device is in clinical use.
!
Warning:
Always wear gloves while cleaning and disinfecting the device.
!
Warning:
Do not service the device while it is in clinical use.
!
Warning:
Always disconnect power prior to performing service or maintenance.
!
Warning:
This device should only be serviced by authorized service personnel.
!
Warning:
Allow the device to dry after cleaning and disinfection. Small amounts of
flammable agents, such as glycol, ether, alcohol, or similar cleaning or disinfecting
solvents left wet on the device can ignite.
!
Warning:
Dispose of all waste properly as per federal, state, and local waste
disposal regulations. Improper disposal could result in personal injury and
environmental impact.
!
Warning:
Using a device with a fan motor that has worn parts can produce
unacceptable noise levels within the patient compartment.
!
Warning:
Limit the load on the device on the outside of each upright to 18kg. Do
not mount shelves to the outside mounting rail more than 137 cm from the floor.
!
Warning:
Overloading the shelves and rails can affect the stability of the device.
Always try to evenly distribute the weight of options on both sides of the unit for a
more balanced load.
Содержание Giraffe OmniBed Carestation CS1
Страница 4: ...RH 2 Service Manual ...
Страница 24: ...Service Language Disclaimer viii Service Manual ...
Страница 40: ...Safety and Regulatory Information User Responsibility xxiv Service Manual ...
Страница 44: ...Chapter 1 Functional Description Control Board 1 2 Service Manual FIGURE 1 1 Block Diagram ...
Страница 74: ...Chapter 3 Calibration and Maintenance Line Voltage Calibration 3 12 Service Manual ...
Страница 82: ...Chapter 4 Understanding the Service Interface Upgrade 4 8 Service Manual ...
Страница 110: ...Chapter 5 Troubleshooting Diagnostic Readings 5 28 Service Manual ...
Страница 211: ...Chapter 7 Repair Procedures Display Harness Connections Service Manual 7 75 ...
Страница 212: ...Chapter 7 Repair Procedures Display Harness Connections 7 76 Service Manual ...
Страница 272: ...Chapter 8 Illustrated Parts Options 8 60 Service Manual FIGURE 8 36 Silo Support Assembly 6600 0853 800 ...
Страница 290: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 78 Service Manual FIGURE 8 45 Wiring Diagram Control Panel ...
Страница 291: ...Chapter 8 Illustrated Parts Wiring Diagrams Service Manual 8 79 FIGURE 8 46 Wiring Diagram Incubator and Warmer Heater ...
Страница 292: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 80 Service Manual FIGURE 8 47 Wiring Diagram Servo Humidifier ...
Страница 293: ...Chapter 8 Illustrated Parts Wiring Diagrams Service Manual 8 81 FIGURE 8 48 Wiring Diagram Servo Control Oxygen ...
Страница 294: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 82 Service Manual ...
Страница 296: ...Appendix A Compartment and Skin Probe Characteristics A 2 Service Manual ...
Страница 304: ...Appendix C RS 232 Serial Data Nurse Call C 4 Service Manual ...