Chapter 2: Installation and Service Checkout
Service Checkout
2-6
Service Manual
Note:
During first 90 minutes of operation, the fan defaults to high speed, and the Air Boost icon is not
operational: .
Note:
Even though the fan is in high speed, the Air Boost icon is not illuminated/animated.
5.
Touch
Air Boost
to illuminate/animate the Air Boost icon (that is, to turn it on).
Note:
The Air Boost icon should now be illuminated/animated:
. The fan remains in high speed.
6.
Touch and hold
Alarm Silence
for 5 to 10 seconds.
The Air Boost icon illumination/animation automatically turns off:
. The fan switches to low fan
speed.
Note:
The Air Boost icon is now fully operational and will increase and decrease the fan speed as desired.
7.
With the fan now in low speed, touch the
Timer
icon and then
Start
to start the timer. Wait 10 minutes
to ensure stability of the Whisper Quiet Mode.
8.
Select
Setup
.
9.
Select
Calibrate Oxygen
on the setup menu to initiate calibration.
10.
When calibration is completed, 100% oxygen is delivered for approximately 20 seconds to ensure there
are no occlusions in the system. Do not turn off the device or disconnect the oxygen supply during this
brief period after the ‘Calibration Complete’ screen appears. Exit calibration screen.
11.
Select O2 icon on display screen to bring up Servo Control Oxygen menu.
12.
Set O2 set point to 65%.
13.
Touch the
Timer
icon then
Start
to start the timer and to verify that device reaches 60% in less than 10
minutes.
Note:
If rise time is longer than 10 minutes, check all tubing between the O2 sensors and the chassis vent
fitting. The chassis vent may be identified by its mushroom cap shaped cover. Make sure that the
fan is in the low speed mode.
Note:
Device should not be placed in service unless the calibration is successful and the rise time test
passes.
Note:
The Low O2 alarm may activate during the rise time test. Silence the alarm and let the test continue.
Note:
The “O2 Cal Lost – No O2” alarm may appear after the device is turned on or returned from power
fail condition. If this alarm appears, calibrate the oxygen sensors (as described in steps 7, 8 & 9) and
resume use. If this alarm appears during normal operation, calibrate and continue use, but call
service.
2.2.4.3 Supply Valve Leak Test
1.
Connect oxygen supply to Servo Oxygen inlet fitting.
2.
Disconnect the 10 mm hose from the expansion chamber.
3.
Power up device in Service Mode. (See “Understanding the Service Interface” on page 4-1.)
Note:
The canopy should be closed for this step through step 9.
4.
Touch
Calibration
in the left-navigation area.
5.
The display area is populated with calibration options.
6.
Touch
Servo O2
.
7.
Open V1 and verify that gas flows audibly.
8.
Close V1 and open V2 and verify that gas flows audibly.
Содержание Giraffe OmniBed Carestation CS1
Страница 4: ...RH 2 Service Manual ...
Страница 24: ...Service Language Disclaimer viii Service Manual ...
Страница 40: ...Safety and Regulatory Information User Responsibility xxiv Service Manual ...
Страница 44: ...Chapter 1 Functional Description Control Board 1 2 Service Manual FIGURE 1 1 Block Diagram ...
Страница 74: ...Chapter 3 Calibration and Maintenance Line Voltage Calibration 3 12 Service Manual ...
Страница 82: ...Chapter 4 Understanding the Service Interface Upgrade 4 8 Service Manual ...
Страница 110: ...Chapter 5 Troubleshooting Diagnostic Readings 5 28 Service Manual ...
Страница 211: ...Chapter 7 Repair Procedures Display Harness Connections Service Manual 7 75 ...
Страница 212: ...Chapter 7 Repair Procedures Display Harness Connections 7 76 Service Manual ...
Страница 272: ...Chapter 8 Illustrated Parts Options 8 60 Service Manual FIGURE 8 36 Silo Support Assembly 6600 0853 800 ...
Страница 290: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 78 Service Manual FIGURE 8 45 Wiring Diagram Control Panel ...
Страница 291: ...Chapter 8 Illustrated Parts Wiring Diagrams Service Manual 8 79 FIGURE 8 46 Wiring Diagram Incubator and Warmer Heater ...
Страница 292: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 80 Service Manual FIGURE 8 47 Wiring Diagram Servo Humidifier ...
Страница 293: ...Chapter 8 Illustrated Parts Wiring Diagrams Service Manual 8 81 FIGURE 8 48 Wiring Diagram Servo Control Oxygen ...
Страница 294: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 82 Service Manual ...
Страница 296: ...Appendix A Compartment and Skin Probe Characteristics A 2 Service Manual ...
Страница 304: ...Appendix C RS 232 Serial Data Nurse Call C 4 Service Manual ...