Chapter 2. Instrument Setup
DMS Go Operator’s Manual
65
2.5.1b Zeroing Out the Probe (PROBE & CAL - PROBE ZERO)
IMPORTANT:
This option is only available if the Zero option (see page 60) in the 1-point calibration mode has been set to Manual or User.
1.
In the
MAIN
menu, scroll to the
PROBE & CAL
submenu using the joystick (
).
2.
Use the joystick (
) to select the
PROBE ZERO
function, then use the joystick (
) to select
Start
.
3.
The menu now varies.
If you have selected Manual in the Zero function, a message appears, “Put probe on zero block -- use
couplant.” Put the probe on the zero block on the DMS Go stand (see the figure at right), apply
couplant to the probe, and take the measurement.
4.
If you have selected User in the Zero option, the message reads, “Wipe off Couplant from Probe, Press
Zero.”
5.
The DMS Go zeroes out the probe. After completing the measurement, use the joystick (
) to
move to another function, or press
HOME
to return to the Main Menu.
Zero
Block