Chapter 2. Instrument Setup
DMS Go Operator’s Manual
37
2.3.2e Setting the Zero Block Configuration (CONFIG-ZERO BLOCK)
If you are using a zero block for your probe, you can enter its thickness and velocity configuration into the DMS Go. (See “Setting the Probe
Zero (PROBE & CAL - ZERO)” on page 63.)
1.
In the
MAIN
menu, scroll to the
CONFIG
submenu using the joystick (
).
2.
Use the joystick (
) to select the
ZERO BLOCK
function, then move the joystick (
) to open the configuration window
Figure 11: Zero Block Configuration Window
3.
Move the joystick (
) to change the unit (inches or mm), thickness (in inches or mm) or the velocity (in inches/microseconds or m/s)
to the value engraved on the GE-supplied reference block. These values will always be displayed in metric (mm). Press F2
(DONE)
to
confirm the entry, or F1
(EXIT)
to close the window without changing the configuration.
4.
After making your choice, use the joystick (
) to move to another function, or press
HOME
to return to the Main Menu.