Chapter 6. Working with Recorded Data
122
DMS Go Operator’s Manual
6.5
Adding a Micro-Grid to a Stored Value
If users wish to record additional thickness measurement data in a single file location, they can create a Micro-Grid. Micro-Grids are small
data subsets within a file. The entire subset is stored at one specific data location. Micro-Grids are created in the Measure mode. They can
only be stored in a file location that already contains a thickness measurement and are placed in the data file location selected at the time of
Micro-Grid creation.
6.5.1
Creating and Navigating Within a Micro-Grid
1.
In the Measure mode, with a data file active and the desired data point (which must already contain a thickness measurement) selected,
center-press and briefly hold the joystick.The function key window will now change, so that Function 1 is
GRID
.
2.
Press Function 1 (
GRID
) The Micro-Grid window opens on the display screen, next to a value representing the grid’s size. Note that
all Micro-Grids must be symmetrical; that is, they must have the same number of rows and columns. Micro-Grids can be created in
sizes ranging from 2 × 2 to 9 × 9.
3.
Set the size of the Micro-Grid by using the joystick to scroll to the desired value.
4.
When you have entered the size information, press F 2 (
DONE
). The DMS Go will return to Measure mode, and the Micro Grid will be
open for data entry. Enter data as you would in a DR file.
5.
When you have completed entering data, press F3 (
GRID
) to close the Micro-Grid.