Chapter 2. Instrument Setup
56
DMS Go Operator’s Manual
2.4.2 Configuring the Instrument
Three instrument settings are directly dependent on the type of probe installed. These settings must be adjusted any time a probe of a
different type is installed, by following the instructions in the following sections.
2.4.2a Determining the Probe Type (PROBE & CAL - PROBE)
1.
Activate the
PROBE & CAL
submenu using the joystick (
).
2.
If the DMS Go is connected to a dialog probe, it automatically detects the probe if the probe selection is on a dual element probe. Once
detected, the screen shows the word “Dialog” below the probe name. If it is not a dialog probe, use the joystick to scroll through the list
of probes and select the appropriate unit. (If you disconnect a dialog probe, then the DMS Go displays the last probe selection, but the
dialog below the probe will no longer appear.) If you change a non-dialog probe, you must again scroll through the list and select the
new unit.
Note:
A dialog probe contains a small chip with probe information. When the probe is connected to the DMS Go, it transmits this
information to the unit.
Note:
The DMS Go offers an option for a GE-supplied custom probe definition file, with a .prb suffix, to be entered via the SD card. The
DMS Go will show this file as part of the Probes list, designated by the (Custom) addition.