Chapter 2. Instrument Setup
DMS Go Operator’s Manual
39
2.3.2f
Setting the Velocity Reference Block Configuration (CONFIG-VEL. REF. BLOCK) (cont.)
1.
In the
MAIN
menu, scroll to the
CONFIG
submenu using the joystick (
).
2.
Use the joystick (
) to select the
VEL. REF. BLOCK
function, and to open the configuration window (Figure 13 below).
Figure 13: Velocity Reference Block Configuration Window
3.
Move the joystick to change the unit measurement (inches or mm), the serial number, the velocity of the base material (in inches or
mm/
sec)
and the coating velocity (the velocity of the coated material, in inches or mm/
sec
) to the value engraved on the block.
Press F2
(DONE)
to confirm the entry, or F1
(EXIT)
to close the window without changing the configuration.
4.
After making your choice, use the joystick (
) to move to another function, or press
HOME
to return to the Main Menu.