Chapter 4. Measuring and Recording Data
86
DMS Go Operator’s Manual
4.1.1c Setting the Gate Threshold (Vertical Position) (GATE - GATE A THR or B THR)
1.
In the
MAIN
menu, scroll to the
GATE
submenu using the joystick (
).
2.
Use the joystick (
) to select the
GATE A THR
(or
B THR
) function.
3.
To change the gate height, either move the joystick (
) or press the function toggle. Increasing or decreasing the value of the
threshold moves the gate up or down, respectively.
4.
After completing your selection, move the joystick up or down (
) to navigate away from this function.
Note:
When RECTIFY is RF, negative threshold values are available.When changed to POS, NEG or FULL, the gate threshold changes to
the corresponding positive values for a short period of time. However, when RECTIFY is switched back to RF, negative positions for
GATE THRESHOLD(s) may be reinstituted again.
Gate B is not permanently available; it is available in TG MODE S-PEAK, S-FLANK for single element probes, and in DUAL
MULTI for some dual element probes.
Your Complete Source for
Testing Equipment. Since 1969!
www.BergEng.com
Berg Engineering & Sales Company, Inc.
1-847-577-3980