Chapter 6. Working with Recorded Data
116
DMS Go Operator’s Manual
6.3.3
Viewing File Statistics (DR-PROPERTIES-STATS) (cont.)
•
Number of data points at the maximum thickness level
•
Number of thickness points above the max-thickness limit
•
Number of points with A-scans attached
•
Number of points with B-scans attached
•
Number of points with Micro-Grids attached
•
The range, mean range, and standard deviation of the data contained in the data file (empty and obstructed points are excluded from
these calculations.)
To view the statistical properties listed above, from the
PROPERTIES
menu, use the joystick to highlight the
STATS
submenu. You can then
scroll to the required statistic. In addition, you can press F2 to
TAG
a specific field to perform a search. For example you can tag “EMPTY
PNTS” . Then you can press F3
(SEARCH)
and the DMS Go will return in the DR section all the empty points. This function aids users to
min, max, empty, and other values.
Note:
The data in this submenu is not directly editable.