5-32
250 Series Maternal/Fetal Monitor
Revision D
2020551-001
Troubleshooting:
FAQs
How do I enable/disable Fetal Alarms
(tachycardia, bradycardia) or Spectra
Alerts?
1. Enter the
Install Options Screen 2
“How do I access the service setup / install
2. In
Install Options Screen 2
, rotate the Trim Knob to highlight the
Fetal Alert/Alarm
:
setting.
3. The current setting is
Off
or
Alarms
. In units where Spectra Alerts have been factory
configured, a third option called
Alerts
appears.
4. Once the
Fetal Alert/Alarm
: setting is highlighted, press the Trim Knob again. The
current setting displays in blinking inverse video.
5. Now rotate the Trim Knob to change the current setting. Select a value.
6. Once you set the desired value, press the Trim Knob again to save the value. The
current value setting stops blinking.
7. To exit the Service mode, follow the instructions at the top of
How do I test the analog outputs from
J102?
1. Enter the
Install Options Screen 1
“How do I access the service setup / install
2. In
Install Options Screen 1
, rotate the Trim Knob to highlight the
Tests
softkey at the
bottom of the display, below the menu bar. This opens the
Diagnostic Control
screen.
3. Rotate the Trim Knob to highlight the
J102
softkey at the bottom of the display, below
the menu bar. This opens the
J102
screen. The
J102
connector is where HR and UA
parameters are available as analog voltages at the rear panel of your monitor.
4. The
J102
screen is divided into five columns:
Pin, Signal, Range, Voltage,
and
Meaning.
Pin
refers to the physical pinout of the connector.
Signal
is the signal schematic designation.
Range
is the expected range of outputs under normal operating conditions.
Voltage
is the actual voltage that should be noted at the indicated
Pin
(using
a DVM referenced to Pin 3, ANALOG GND) while running these tests.
Meaning
is intended interpretation of the voltage indicated in the
Voltage
field
to an external device.
5. To test these outputs, rotate the Trim Knob to the desired row (
Pin
). The cursor can
only move up and down through the last column.
6. Once the desired row is selected, press the Trim Knob. The current setting
(
Meaning
) displays in blinking inverse video.
7. Now rotate the Trim Knob. As the Trim Knob rotates, the value in the
Meaning
column change, as will the voltage indicated in the
Voltage
column. This same
voltage should be seen at the J102 Connector Pin.
8. Once the desired value / pinout is tested and calibrated, press the Trim Knob again
to end the value test. The current value setting stops blinking.
9. Repeat Steps 5 through 8, as required.
10. To exit the Service mode, follow the instructions at the top of
Question
Answer
Содержание Corometrics 250 Series
Страница 2: ......
Страница 6: ...CE CE ii 0086 ...
Страница 14: ...viii 250 Series Maternal Fetal Monitor Revision D 2020551 001 ...
Страница 16: ...1 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 28: ...1 14 250 Series Maternal Fetal Monitor Revision D 2020551 001 Introduction Service Requirements ...
Страница 29: ...Revision D 250 Series Maternal Fetal Monitor 2 1 2020551 001 2 Equipment Overview ...
Страница 30: ...2 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 54: ...2 26 250 Series Maternal Fetal Monitor Revision D 2020551 001 Equipment Overview Theory of Operation ...
Страница 55: ...3 1 250 Series Maternal Fetal Monitor Revision D 2020551 001 3 Installation ...
Страница 56: ...3 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 81: ...Revision D 250 Series Maternal Fetal Monitor 4 1 2020551 001 4 Maintenance ...
Страница 82: ...4 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 142: ...4 62 250 Series Maternal Fetal Monitor Revision D 2020551 001 Maintenance Preventative Maintenance Inspection Report ...
Страница 143: ...Revision D 250 Series Maternal Fetal Monitor 5 1 2020551 001 5 Troubleshooting ...
Страница 144: ...5 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 194: ...5 52 250 Series Maternal Fetal Monitor Revision D 2020551 001 Troubleshooting General Troubleshooting ...
Страница 195: ...Revision D 250 Series Maternal Fetal Monitor 6 1 2020551 001 6 Parts List Drawings and Replacement ...
Страница 196: ...6 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 228: ...A 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 Technical Specifications For your notes ...
Страница 238: ...A 12 250 Series Maternal Fetal Monitor Revision D 2020551 001 Technical Specifications Strip Chart Recorder ...
Страница 239: ...Revision D 250 Series Maternal Fetal Monitor B 1 2020551 001 B Alarms Summary ...
Страница 240: ...B 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 Alarms Summary For your notes ...
Страница 243: ...Revision D 250 Series Maternal Fetal Monitor C 1 2020551 001 C Electromagnetic Compatibility ...
Страница 244: ...C 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 Electromagnetic Compatibility For your notes ...
Страница 251: ......