5-34
250 Series Maternal/Fetal Monitor
Revision D
2020551-001
Troubleshooting:
FAQs
How do I test the RS232 outputs?
1. The 250 Series Monitor Serial Interface is capable of being tested by means of a
loop back connector. A loop back connector simply routes the TX line from the serial
port to the RX line of the same serial port.
2. Enter the
Install Options Screen 1
“How do I access the service setup / install
3. In the
Install Options Screen 1
, rotate the Trim Knob to highlight the
COMM
softkey
at the bottom of the display below the menu bar.
4. Press the Trim Knob to enter the
Communications Setup
screen.
5. On the
Communications Setup
screen, the three items in the column at the far left
represent the rear panel RS232 Communications Port designations
J109, J110
and
J111.
The middle column represents the current
Baudrate
for the associated
communications port (same line reading across), and the last column on the right
represents the communication protocol for the associated communication port.
6. To change a
Baudrate
, rotate the Trim Knob to highlight the desired baudrate
selection on the desired communication port line.
7. Press the Trim Knob. The current setting displays in blinking inverse video.
8. Now rotate the Trim Knob to change the current baudrate setting. Select from one of
the following:
600, 1200, 2400, 4800, 9600,
and
19200.
9. When the desired test baudrate displays, press the Trim Knob again to save. The
current value stops blinking.
10. Note the current protocol setting so that it may be reset when testing is completed.
11. Then rotate the Trim Knob to highlight the protocol selection of the desired
communication port line.
12. Press the Trim Knob. The current setting displays in blinking inverse video.
13. Now rotate the Trim Knob to change the current setting to
Loopback
.
14. The column immediately to the right now displays a new indication,
Off
.
15. Insert the loopback connector in the corresponding rear panel serial port connector.
Within a few seconds the
Off
indication should change to
OK
.
16. Return the protocol setting to the setting previously noted, and press the Trim Knob
to save. The current value stops blinking.
17. Repeat Steps 6 through 16 for each of the serial ports.
18. To exit the Service mode following the instructions at the top of
Question
Answer
Содержание Corometrics 250 Series
Страница 2: ......
Страница 6: ...CE CE ii 0086 ...
Страница 14: ...viii 250 Series Maternal Fetal Monitor Revision D 2020551 001 ...
Страница 16: ...1 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 28: ...1 14 250 Series Maternal Fetal Monitor Revision D 2020551 001 Introduction Service Requirements ...
Страница 29: ...Revision D 250 Series Maternal Fetal Monitor 2 1 2020551 001 2 Equipment Overview ...
Страница 30: ...2 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 54: ...2 26 250 Series Maternal Fetal Monitor Revision D 2020551 001 Equipment Overview Theory of Operation ...
Страница 55: ...3 1 250 Series Maternal Fetal Monitor Revision D 2020551 001 3 Installation ...
Страница 56: ...3 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 81: ...Revision D 250 Series Maternal Fetal Monitor 4 1 2020551 001 4 Maintenance ...
Страница 82: ...4 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 142: ...4 62 250 Series Maternal Fetal Monitor Revision D 2020551 001 Maintenance Preventative Maintenance Inspection Report ...
Страница 143: ...Revision D 250 Series Maternal Fetal Monitor 5 1 2020551 001 5 Troubleshooting ...
Страница 144: ...5 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 194: ...5 52 250 Series Maternal Fetal Monitor Revision D 2020551 001 Troubleshooting General Troubleshooting ...
Страница 195: ...Revision D 250 Series Maternal Fetal Monitor 6 1 2020551 001 6 Parts List Drawings and Replacement ...
Страница 196: ...6 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 228: ...A 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 Technical Specifications For your notes ...
Страница 238: ...A 12 250 Series Maternal Fetal Monitor Revision D 2020551 001 Technical Specifications Strip Chart Recorder ...
Страница 239: ...Revision D 250 Series Maternal Fetal Monitor B 1 2020551 001 B Alarms Summary ...
Страница 240: ...B 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 Alarms Summary For your notes ...
Страница 243: ...Revision D 250 Series Maternal Fetal Monitor C 1 2020551 001 C Electromagnetic Compatibility ...
Страница 244: ...C 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 Electromagnetic Compatibility For your notes ...
Страница 251: ......