4-36
250 Series Maternal/Fetal Monitor
Revision D
2020551-001
Maintenance:
Strain Gauge Transducer Test
4. Momentarily depress the monitor’s
UA Reference
button. Verify the following:
The UA value shows the
default
setting.
The UA mode shows
TOCO
.
The recorder prints the messages
UA REF
and
TOCO
on the strip chart paper.
5. Apply gentle pressure to the tocotransducer diaphragm and verify that the UA
value responds to the pressure input. Increasing force should produce an
increasing value and vice versa.
6. Remove the tocotransducer from the monitor’s
UA
input connector. Verify the
following on the monitor:
The UA value and INOP are both blank.
The recorder stops printing the uterine activity trace.
The recorder prints the message
UA INOP
on the center margin of the strip
chart paper after approximately 20 seconds.
Strain Gauge Transducer Test
1. Inspect a strain gauge as follows:
Unscrew the plastic dome from the transducer and check for any cracks or
contaminants on the transducer.
Visibly assess the condition of the cable, strain relief, and the connector
pins.
2. Connect the strain gauge to the
UA
input connector on the front panel of the 250
Series Monitor. Verify the following on the monitor:
The UA value may read negative numbers indicating baseline pressure is
off scale. In this case, the recorder prints the message
BASELINE PRESSURE
OFFSCALE
on the bottom of the
UA
scale on the strip chart paper.
The UA mode is
IUP
for intrauterine pressure.
The recorder prints the message
IUP
on the center margin of the strip chart
paper after approximately 20 seconds.
3. Depress the monitor’s
UA Reference
button and verify the following on the
monitor:
The UA value is 0 mmHg (0 kPa).
The recorder prints a continuous line at 0 mmHg (0 kPa) on the strip chart
paper.
The recorder prints the message
UA REF
on the bottom two lines of the top
grid of the strip chart paper.
4. Apply gentle pressure on the strain gauge diaphragm and verify that the display
and recorder respond to the input. Increasing force should produce an
increasing value and vice versa.
5. Disconnect the strain gauge from the front panel of the monitor. Verify the
following on the monitor:
Содержание Corometrics 250 Series
Страница 2: ......
Страница 6: ...CE CE ii 0086 ...
Страница 14: ...viii 250 Series Maternal Fetal Monitor Revision D 2020551 001 ...
Страница 16: ...1 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 28: ...1 14 250 Series Maternal Fetal Monitor Revision D 2020551 001 Introduction Service Requirements ...
Страница 29: ...Revision D 250 Series Maternal Fetal Monitor 2 1 2020551 001 2 Equipment Overview ...
Страница 30: ...2 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 54: ...2 26 250 Series Maternal Fetal Monitor Revision D 2020551 001 Equipment Overview Theory of Operation ...
Страница 55: ...3 1 250 Series Maternal Fetal Monitor Revision D 2020551 001 3 Installation ...
Страница 56: ...3 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 81: ...Revision D 250 Series Maternal Fetal Monitor 4 1 2020551 001 4 Maintenance ...
Страница 82: ...4 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 142: ...4 62 250 Series Maternal Fetal Monitor Revision D 2020551 001 Maintenance Preventative Maintenance Inspection Report ...
Страница 143: ...Revision D 250 Series Maternal Fetal Monitor 5 1 2020551 001 5 Troubleshooting ...
Страница 144: ...5 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 194: ...5 52 250 Series Maternal Fetal Monitor Revision D 2020551 001 Troubleshooting General Troubleshooting ...
Страница 195: ...Revision D 250 Series Maternal Fetal Monitor 6 1 2020551 001 6 Parts List Drawings and Replacement ...
Страница 196: ...6 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 228: ...A 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 Technical Specifications For your notes ...
Страница 238: ...A 12 250 Series Maternal Fetal Monitor Revision D 2020551 001 Technical Specifications Strip Chart Recorder ...
Страница 239: ...Revision D 250 Series Maternal Fetal Monitor B 1 2020551 001 B Alarms Summary ...
Страница 240: ...B 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 Alarms Summary For your notes ...
Страница 243: ...Revision D 250 Series Maternal Fetal Monitor C 1 2020551 001 C Electromagnetic Compatibility ...
Страница 244: ...C 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 Electromagnetic Compatibility For your notes ...
Страница 251: ......