Revision D
250 Series Maternal/Fetal Monitor
3-19
2020551-001
Installation:
Service Mode Screens
NIBP 1 min Interval
This field is used to enable/disable the 1 minute interval selection on the maternal
BP Setup screen. Use the Trim Knob control to alternate between
On
and
Off
. (The
factory default setting is
Off
.)
NIBP Display
Setting this field to a value reduces the chance of error. Setting this field to
On
leaves
the blood pressure reading displayed indefinitely and could potentially cause
confusion. For example: if the monitor is configured for manual mode and 1 hour
has elapsed since the last reading, the continuous display of the “old” NIBP reading
may cause confusion.
This field determines the time period, in minutes, that a blood pressure reading
remains displayed before being automatically erased,
1
starting from the time the
reading is displayed.
Pressure Units
The
Pressure Units
allows you choose either mmHg or kPa.
SatSeconds
The
SatSeconds
limit controls the time that the SpO
2
level may fall outside the alarm
before an audible alarm sounds. The
SatSeconds
default setting is set at 10
SatSeconds.
Your biomed may set the
SatSeconds
limit (
Off
,
10, 25, 50,
or
100
) to suit
the clinical environment and patient condition.
Default Settings
The 250 Series Monitor is shipped with factory defaults for the setup screens. You
can change most of these settings to suit your particular needs. The monitor has two
Default Settings:
that allow you to determine the settings on the next power-on or
restart:
Factory
or
Hospital
.
Factory:
Select this option if you wish to return all monitor settings back to the
factory default settings (refer to the table at the end of this section). Factory
defaults represent settings chosen to meet the requirements of a typical labor
and delivery environment.
NOTE
:
You may adjust the setup screen configurations as needed during
monitoring; however, be advised that if you turn off the monitor, all
user setup screens revert to the factory defaults when the monitor is
turned on again.
Hospital:
The
Hospital
option allows your hospital to configure its own set of
preferences to be used each time the monitor is powered on. Record your
hospital settings in the table at the end of this section. You may adjust the setup
screen configurations as needed during monitoring; however, be advised that if
you turn off the monitor, all user setup screens revert to the factory defaults
when the monitor is turned on again.
Содержание Corometrics 250 Series
Страница 2: ......
Страница 6: ...CE CE ii 0086 ...
Страница 14: ...viii 250 Series Maternal Fetal Monitor Revision D 2020551 001 ...
Страница 16: ...1 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 28: ...1 14 250 Series Maternal Fetal Monitor Revision D 2020551 001 Introduction Service Requirements ...
Страница 29: ...Revision D 250 Series Maternal Fetal Monitor 2 1 2020551 001 2 Equipment Overview ...
Страница 30: ...2 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 54: ...2 26 250 Series Maternal Fetal Monitor Revision D 2020551 001 Equipment Overview Theory of Operation ...
Страница 55: ...3 1 250 Series Maternal Fetal Monitor Revision D 2020551 001 3 Installation ...
Страница 56: ...3 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 81: ...Revision D 250 Series Maternal Fetal Monitor 4 1 2020551 001 4 Maintenance ...
Страница 82: ...4 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 142: ...4 62 250 Series Maternal Fetal Monitor Revision D 2020551 001 Maintenance Preventative Maintenance Inspection Report ...
Страница 143: ...Revision D 250 Series Maternal Fetal Monitor 5 1 2020551 001 5 Troubleshooting ...
Страница 144: ...5 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 194: ...5 52 250 Series Maternal Fetal Monitor Revision D 2020551 001 Troubleshooting General Troubleshooting ...
Страница 195: ...Revision D 250 Series Maternal Fetal Monitor 6 1 2020551 001 6 Parts List Drawings and Replacement ...
Страница 196: ...6 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 228: ...A 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 Technical Specifications For your notes ...
Страница 238: ...A 12 250 Series Maternal Fetal Monitor Revision D 2020551 001 Technical Specifications Strip Chart Recorder ...
Страница 239: ...Revision D 250 Series Maternal Fetal Monitor B 1 2020551 001 B Alarms Summary ...
Страница 240: ...B 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 Alarms Summary For your notes ...
Страница 243: ...Revision D 250 Series Maternal Fetal Monitor C 1 2020551 001 C Electromagnetic Compatibility ...
Страница 244: ...C 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 Electromagnetic Compatibility For your notes ...
Страница 251: ......