Revision D
250 Series Maternal/Fetal Monitor
4-51
2020551-001
Maintenance:
RS-232C Connector Loopback Test
RS-232C Connector Loopback Test
The Communications Setup screen includes a loopback test option for each of the
communications ports:
J109, J110,
and
J111.
Running the test requires inserting a
loopback test connector into each port being tested. (You can test more than one port
at a time.)
Making a Loopback Test Connector
Start with an RJ-11C connector and bridge pin 2 (RXD) and pin 5 (TXD) together.
The figure below shows the pinout of a communications port while you face the rear
panel of a 250 Series Monitor.
lists the signal description for each pin.
Testing the Port(s)
1. Access the
Communications Setup
screen (See
2. Insert a loopback test connector into each communications port being tested.
3. Set the mode field on the
Communications Setup
screen to
Loopback
. Notice that
the word
Off
displays to the right of the mode.
4. Verify that, after a few seconds, the status
Loopback
OK
displays.
OK
indicates
that the test has passed.
NOTE:
If
Off
remains displayed, the test has failed; contact your GE Service
Representative. Do not use this port to connect to any peripheral
equipment until the port or Communications Board has been serviced.
RJ-11C Connector
Table 18. DAC Output Voltages and Adjustment Points for Hewlett-Packard Central Station Outputs
Analog Signal
Measurement Site
Service Screen
Setting
Expected Voltage
Adjustment Site
HR1
J102, pin 7
240 bpm
+2.4V ± 10 mV
R20
HR2
J102, pin 22
240 bpm
+2.4 V ± 10 mV
R29
UA
J102, pin 2
100 mmHg
13.3 kPa
+10.0 V ± 24 mV
R44
Содержание Corometrics 250 Series
Страница 2: ......
Страница 6: ...CE CE ii 0086 ...
Страница 14: ...viii 250 Series Maternal Fetal Monitor Revision D 2020551 001 ...
Страница 16: ...1 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 28: ...1 14 250 Series Maternal Fetal Monitor Revision D 2020551 001 Introduction Service Requirements ...
Страница 29: ...Revision D 250 Series Maternal Fetal Monitor 2 1 2020551 001 2 Equipment Overview ...
Страница 30: ...2 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 54: ...2 26 250 Series Maternal Fetal Monitor Revision D 2020551 001 Equipment Overview Theory of Operation ...
Страница 55: ...3 1 250 Series Maternal Fetal Monitor Revision D 2020551 001 3 Installation ...
Страница 56: ...3 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 81: ...Revision D 250 Series Maternal Fetal Monitor 4 1 2020551 001 4 Maintenance ...
Страница 82: ...4 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 142: ...4 62 250 Series Maternal Fetal Monitor Revision D 2020551 001 Maintenance Preventative Maintenance Inspection Report ...
Страница 143: ...Revision D 250 Series Maternal Fetal Monitor 5 1 2020551 001 5 Troubleshooting ...
Страница 144: ...5 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 194: ...5 52 250 Series Maternal Fetal Monitor Revision D 2020551 001 Troubleshooting General Troubleshooting ...
Страница 195: ...Revision D 250 Series Maternal Fetal Monitor 6 1 2020551 001 6 Parts List Drawings and Replacement ...
Страница 196: ...6 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 228: ...A 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 Technical Specifications For your notes ...
Страница 238: ...A 12 250 Series Maternal Fetal Monitor Revision D 2020551 001 Technical Specifications Strip Chart Recorder ...
Страница 239: ...Revision D 250 Series Maternal Fetal Monitor B 1 2020551 001 B Alarms Summary ...
Страница 240: ...B 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 Alarms Summary For your notes ...
Страница 243: ...Revision D 250 Series Maternal Fetal Monitor C 1 2020551 001 C Electromagnetic Compatibility ...
Страница 244: ...C 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 Electromagnetic Compatibility For your notes ...
Страница 251: ......