5-30
250 Series Maternal/Fetal Monitor
Revision D
2020551-001
Troubleshooting:
FAQs
How do I change the default
tocodynamometer reference value?
1. Enter
Install Options Screen 2
. (Refer to
“How do I access the service setup / install
2. In
Install Options Screen 2
, rotate the Trim Knob to highlight the
Default TOCO
Reference
: setting on the eighth line from the top (excluding the Title).
3. The current setting is one of the following:
5, 10, 15, 20
or
25 mmHg.
4. Once the
Default TOCO Reference
: setting is highlighted, press the Trim Knob. The
current setting displays in blinking inverse video.
5. Now rotate the Trim Knob to change the current setting. Select a value.
6. Once the desired value is set, press the Trim Knob again to save the value. The
current value setting stops blinking.
7. To exit the Service mode, follow the instructions at the top of
How do I change the date?
1. Rotate the Trim Knob to highlight the
Setup
softkey at the bottom of the display,
below the menu bar.
2. Once the
Setup
softkey is highlighted, press the Trim Knob.
3. The display will change to the
General
Setup
screen.
4. In the
General
Setup
screen, rotate the Trim Knob to highlight one of the
Date:
setting fields on the top right corner of the display. These fields are for day (DD),
month (MMM), and year (YYYY).
5. After the desired field is highlighted (selected), press the Trim Knob. The current
setting displays in blinking inverse video.
6. Now rotate the Trim Knob to change the current date parameter setting.
7. Once you set the desired value, press the Trim Knob again to save the value. The
current value setting stops blinking.
8. Repeat Step 4 through Step 7 for any other date parameters that need to be set.
9. To exit the Service mode, follow the instructions at the top of
How do I change the time?
1. Rotate the Trim Knob to highlight the
Setup
softkey at the bottom of the display,
below the menu bar.
2. Once the
Setup
softkey is highlighted, press the Trim Knob.
3. The display changes to the
General
Setup
screen.
4. In the
General
Setup
screen, rotate the Trim Knob to highlight one of the
Time:
setting fields on the top left corner of the display. These fields are for hours (HH),
minutes (MMM), and seconds (YYYY). Note that the seconds field cannot be
selected or set.
5. After the desired field is highlighted (selected), press the Trim Knob. The current
setting displays in blinking inverse video.
6. Now rotate the Trim Knob to change the current time parameter setting.
7. Once you set the desired value, press the Trim Knob again to save the value. The
current value setting stops blinking.
8. Repeat Step 4 through Step 7 for any other date parameters that need to be set.
9. To exit the Service mode, follow the instructions at the top of
Question
Answer
Содержание Corometrics 250 Series
Страница 2: ......
Страница 6: ...CE CE ii 0086 ...
Страница 14: ...viii 250 Series Maternal Fetal Monitor Revision D 2020551 001 ...
Страница 16: ...1 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 28: ...1 14 250 Series Maternal Fetal Monitor Revision D 2020551 001 Introduction Service Requirements ...
Страница 29: ...Revision D 250 Series Maternal Fetal Monitor 2 1 2020551 001 2 Equipment Overview ...
Страница 30: ...2 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 54: ...2 26 250 Series Maternal Fetal Monitor Revision D 2020551 001 Equipment Overview Theory of Operation ...
Страница 55: ...3 1 250 Series Maternal Fetal Monitor Revision D 2020551 001 3 Installation ...
Страница 56: ...3 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 81: ...Revision D 250 Series Maternal Fetal Monitor 4 1 2020551 001 4 Maintenance ...
Страница 82: ...4 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 142: ...4 62 250 Series Maternal Fetal Monitor Revision D 2020551 001 Maintenance Preventative Maintenance Inspection Report ...
Страница 143: ...Revision D 250 Series Maternal Fetal Monitor 5 1 2020551 001 5 Troubleshooting ...
Страница 144: ...5 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 194: ...5 52 250 Series Maternal Fetal Monitor Revision D 2020551 001 Troubleshooting General Troubleshooting ...
Страница 195: ...Revision D 250 Series Maternal Fetal Monitor 6 1 2020551 001 6 Parts List Drawings and Replacement ...
Страница 196: ...6 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 228: ...A 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 Technical Specifications For your notes ...
Страница 238: ...A 12 250 Series Maternal Fetal Monitor Revision D 2020551 001 Technical Specifications Strip Chart Recorder ...
Страница 239: ...Revision D 250 Series Maternal Fetal Monitor B 1 2020551 001 B Alarms Summary ...
Страница 240: ...B 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 Alarms Summary For your notes ...
Страница 243: ...Revision D 250 Series Maternal Fetal Monitor C 1 2020551 001 C Electromagnetic Compatibility ...
Страница 244: ...C 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 Electromagnetic Compatibility For your notes ...
Страница 251: ......