Revision D
250 Series Maternal/Fetal Monitor
4-59
2020551-001
Maintenance:
Preventative Maintenance Inspection Report
Page
Description
Complete
Max
Actual
Pass
Fail
N/A
Cleaning
Clean the monitor exterior.
Clean the monitor interior using a hand-held
vacuum.
Clean the recorder printhead.
Clean the UA Strain Gauge.
Clean the monitor accessories (TOCO,
transducers, cables)
Clean the monitor accessories (cuffs, hoses)
Calibration
Page
Description
Complete
Max
Actual
Pass
Fail
N/A
Verify the power supply voltages on the Main
Board.
Verify the power supply voltages on the Isolated
Power Supply Board.
Adjust the printhead.
Adjust the photosensors on the Recorder Board.
Electrical Safety Tests
Page
Description
Complete
Max
Actual
Pass
Fail
N/A
AC Line (120 V
10%)
(<300 mW)
Ground Impedance (<300 M
Ω
)
Lead to rear panel ground lug (<300 M
Ω
)
Lead to top cover bezel screw (<300 M
Ω
)
(<300 mW)
Unit to primary leakage
(<300
mA)
Patient-to-Ground leakage for ECG
(
<
50mA)
Patient-to-Line leakage for ECG
(
<
50mA)
Patient-to-Ground leakage for IUP
(
<
50mA)
Patient-to-line leakage for IUP
(
<
50mA)
Patient-to-Ground leakage for MSpO
2
(
<
50mA)
Patient-to-Line leakage for MSpO
2
(
<
50mA)
Patient-to-Ground leakage for US
(
<
50mA)
↑
Содержание Corometrics 250 Series
Страница 2: ......
Страница 6: ...CE CE ii 0086 ...
Страница 14: ...viii 250 Series Maternal Fetal Monitor Revision D 2020551 001 ...
Страница 16: ...1 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 28: ...1 14 250 Series Maternal Fetal Monitor Revision D 2020551 001 Introduction Service Requirements ...
Страница 29: ...Revision D 250 Series Maternal Fetal Monitor 2 1 2020551 001 2 Equipment Overview ...
Страница 30: ...2 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 54: ...2 26 250 Series Maternal Fetal Monitor Revision D 2020551 001 Equipment Overview Theory of Operation ...
Страница 55: ...3 1 250 Series Maternal Fetal Monitor Revision D 2020551 001 3 Installation ...
Страница 56: ...3 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 81: ...Revision D 250 Series Maternal Fetal Monitor 4 1 2020551 001 4 Maintenance ...
Страница 82: ...4 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 142: ...4 62 250 Series Maternal Fetal Monitor Revision D 2020551 001 Maintenance Preventative Maintenance Inspection Report ...
Страница 143: ...Revision D 250 Series Maternal Fetal Monitor 5 1 2020551 001 5 Troubleshooting ...
Страница 144: ...5 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 194: ...5 52 250 Series Maternal Fetal Monitor Revision D 2020551 001 Troubleshooting General Troubleshooting ...
Страница 195: ...Revision D 250 Series Maternal Fetal Monitor 6 1 2020551 001 6 Parts List Drawings and Replacement ...
Страница 196: ...6 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 228: ...A 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 Technical Specifications For your notes ...
Страница 238: ...A 12 250 Series Maternal Fetal Monitor Revision D 2020551 001 Technical Specifications Strip Chart Recorder ...
Страница 239: ...Revision D 250 Series Maternal Fetal Monitor B 1 2020551 001 B Alarms Summary ...
Страница 240: ...B 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 Alarms Summary For your notes ...
Страница 243: ...Revision D 250 Series Maternal Fetal Monitor C 1 2020551 001 C Electromagnetic Compatibility ...
Страница 244: ...C 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 Electromagnetic Compatibility For your notes ...
Страница 251: ......