Revision D
250 Series Maternal/Fetal Monitor
5-29
2020551-001
Troubleshooting:
FAQs
How do I enable/disable
Smart BP?
1. Enter
Install Options Screen 1
“How do I access the service setup / install
2. In
Install Options Screen 1
, rotate the Trim Knob to highlight the
Smart BP
: setting.
3. The current setting is ON (enabled) or
Off
(disabled).
4. Once the
Smart BP
: setting is highlighted, press the Trim Knob again. The current
setting displays in blinking inverse video.
5. Now rotate the Trim Knob to change the current volume setting. Select a value.
6. Once you set the desired value, press the Trim Knob again to save the value. The
current value setting stops blinking.
7. To exit the Service mode, follow the instructions at the top of
How do I change the interval time for taking
Non-Invasive Blood Pressures?
1. Rotate the Trim Knob to highlight the field for
NIBP
. (This field is slightly above
center, on the left side of the display.)
2. Once the
NIBP
field is highlighted, press the Trim Knob. The display changes to
show the
NIBP Setup
screen.
3. Now rotate the Trim Knob to highlight the
Mode
: setting which is below the
NIBP
Setup
title. The current setting is one of the following:
Manual, Auto 1 min, Auto 2
min, Auto 3 min, Auto 4 min, Auto 5 min, Auto 10 min, Auto 20 min, Auto 30 min,
Auto 40 min, Auto 45 min, Auto 60 min, Auto 90 min, or Auto 120 min.
4. Once the
Mode
: setting is highlighted, press the Trim Knob again. The current setting
displays in blinking inverse video.
5. Now rotate the Trim Knob to change the current setting. Select a value from one of
the following:
Manual, Auto 1 min, Auto 2 min, Auto 3 min, Auto 4 min, Auto 5 min,
Auto 10 min, Auto 20 min, Auto 30 min, Auto 40 min, Auto 45 min, Auto 60 min, Auto
90 min, or Auto 120 min.
6. Once you set the desired alarm value, press the Trim Knob to confirm your selection.
The current value setting stops blinking.
7. To exit the Service mode, follow the instructions at the top of
How do I change the display time-out
duration for Non-Invasive Blood
Pressures?
1. Navigate to
Install Options Screen 2
.
2. Now rotate the Trim Knob to highlight the
NIBP Display
: setting. The current setting
is one of the following:
On
(do not blank last NIBP taken),
1 min Blank
(blank last
NIBP taken after 1 minute),
2 min Blank, 3 min Blank, 5 min Blank, 10 min Blank, 15
min Blank, 30 min Blank.
3. Once the
Display
: setting is highlighted, press the Trim Knob again. The current
setting displays in blinking inverse video.
4. Now rotate the Trim Knob to change the current setting. Select one of the following:
On
,
1 min Blank, 2 min Blank, 3 min Blank, 5 min Blank, 10 min Blank, 15 min Blank,
30 min Blank.
5. Once you set the desired
NIBP Display
: value, press the Trim Knob again to save the
value. The current value setting stops blinking.
6. To exit the Service mode, follow the instructions at the top of
How do I activate and terminate Heart Rate
Offset? Now in
Install Options Screen 2
1. Press and hold the
Mark [Offset]
button (over the recorder) on the front panel of the
250 Series Monitor for 2 seconds. You will hear a short, two-tone audio beep from
the monitor when offset is activated.
2. To end the HR offset period, press and hold the
Mark [Offset]
button (over the
recorder) again for 2 seconds. You will hear a short, two-tone audio beep from the
monitor when
HR Offset
is
Off
.
How do I know when the monitor detects
Heart Beat Coincidence?
Heart Beat Coincidence (
HBC
) indicates that two HR channels may be monitoring the
same signal by placing both heart rates in inverse video on the front panel display, AND
by placing a symbol of two overlapping hearts on the trend recorder.
Question
Answer
Содержание Corometrics 250 Series
Страница 2: ......
Страница 6: ...CE CE ii 0086 ...
Страница 14: ...viii 250 Series Maternal Fetal Monitor Revision D 2020551 001 ...
Страница 16: ...1 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 28: ...1 14 250 Series Maternal Fetal Monitor Revision D 2020551 001 Introduction Service Requirements ...
Страница 29: ...Revision D 250 Series Maternal Fetal Monitor 2 1 2020551 001 2 Equipment Overview ...
Страница 30: ...2 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 54: ...2 26 250 Series Maternal Fetal Monitor Revision D 2020551 001 Equipment Overview Theory of Operation ...
Страница 55: ...3 1 250 Series Maternal Fetal Monitor Revision D 2020551 001 3 Installation ...
Страница 56: ...3 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 81: ...Revision D 250 Series Maternal Fetal Monitor 4 1 2020551 001 4 Maintenance ...
Страница 82: ...4 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 142: ...4 62 250 Series Maternal Fetal Monitor Revision D 2020551 001 Maintenance Preventative Maintenance Inspection Report ...
Страница 143: ...Revision D 250 Series Maternal Fetal Monitor 5 1 2020551 001 5 Troubleshooting ...
Страница 144: ...5 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 194: ...5 52 250 Series Maternal Fetal Monitor Revision D 2020551 001 Troubleshooting General Troubleshooting ...
Страница 195: ...Revision D 250 Series Maternal Fetal Monitor 6 1 2020551 001 6 Parts List Drawings and Replacement ...
Страница 196: ...6 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Страница 228: ...A 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 Technical Specifications For your notes ...
Страница 238: ...A 12 250 Series Maternal Fetal Monitor Revision D 2020551 001 Technical Specifications Strip Chart Recorder ...
Страница 239: ...Revision D 250 Series Maternal Fetal Monitor B 1 2020551 001 B Alarms Summary ...
Страница 240: ...B 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 Alarms Summary For your notes ...
Страница 243: ...Revision D 250 Series Maternal Fetal Monitor C 1 2020551 001 C Electromagnetic Compatibility ...
Страница 244: ...C 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 Electromagnetic Compatibility For your notes ...
Страница 251: ......