5
Troubleshooting
TRM, Centiva/5 Plus, M1047448/CM-922026, rev. 3.50.a of 11/05
5-13
Code Message
Type
Set condition
Action
Reset condition
09-2 EPROM
on
front PCB
defective.
Repeat test
by turning
OFF and ON
again.
System shut
down error
EPROM problems are
identified on panel PC board
Check proper seating of
EPROM in socket and
replace A EPROM, if
necessary
EPROM is identified to
be OK.
09-3 EEPROM
on
main PCB
defective.
Repeat test
by turning
OFF and ON
again.
System shut
down error
EEPROM problems are
identified on main PC board
Repeat system test for 5
times to load new data into
EEPROM or replace B
board, if necessary
EEPROM is identified to
be OK.
09-4 EEPROM
on
front PCB
defective.
Repeat test
by turning
OFF and ON
again.
System shut
down error
EEPROM problems are
identified on panel PC board
Repeat system test for 5
times to load new data into
EEPROM or replace A
board, if necessary
EEPROM is identified to
be OK.
10-1 Button
defective or
permanently
pressed.
Repeat test
by turning
OFF and ON
again.
Advice
Any front panel button is
defective
Or
Any button is permanently
activated
Check button connection
within control panel, if
necessary replace front
label with buttons
Check for permanently
activated button on
front panel
Or repeat test by
turning OFF and ON
again
10-2 Buzzer
on
main PCB
defective.
Repeat test
by turning
OFF and ON
again.
System shut
down error
Buzzer problems are
identified on main PC board
Check connection PC
board B and Alarm system,
if necessary replace main
alarm system
Buzzer is identified to
be OK.
10-3 Buzzer
on
front PCB
defective.
Repeat test
by turning ON
and Off
again.
System shut
down error
Buzzer problems are
identified on panel PC board
Check connection PC
board A and Alarm system,
if necessary replace front
alarm system
Buzzer is identified to
be OK.
Содержание Centiva/5 Plus
Страница 1: ...GE Healthcare Centiva 5 Plus Technical Reference Manual ...
Страница 14: ...1 Introduction TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 1 8 Notes ...
Страница 20: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 6 ...
Страница 27: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 13 ...
Страница 30: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 16 Notes ...
Страница 34: ...3 Post service check out TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 3 4 Notes ...
Страница 70: ...4 Test and calibration TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 4 36 Notes ...
Страница 154: ...7 Repair procedures TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a 7 48 Notes ...
Страница 178: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 2 9 1 Pneumatic diagram ...
Страница 180: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 4 9 2 Electrical diagram ...
Страница 181: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 5 ...
Страница 182: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 6 9 3 Component layout PC board B ...
Страница 183: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 7 9 4 Component layout PC board A ...
Страница 192: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 16 Notes ...
Страница 206: ...10 Configuration and optionalization TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 10 14 Notes ...
Страница 212: ...11 Optional accessories TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 11 6 Notes ...