9 Schematics and Diagrams
TRM, Centiva/5 Plus, M1047448/CM-922026, rev. 3.50.a of 11/05
9-14
The following data ranges deal with the four pressure sensors, used to measure
•
Pat. pressure = Airway pressure
•
Exp. flow = Expiration flow, measured via differential pressure
•
AIR flow = Inspiration flow AIR, measured via differential pressure
•
O2 flow = Inspiration flow O2, measured via differential pressure
The sensor offset values are defined in different accepted ranges
•
No readjust range = Offset value is within an area where re-
adjustment is not necessary.
•
Max. auto correct range = Offset values are automatically
corrected by the software during auto zeroing
For sensor offset values that are outside the “No readjust” offset re-adjustment is
necessary.
According to the sensor’s warm-up drift always allow the Centiva/5 to warm up for
at least 15 minutes, before taking data or performing any calibration.
Rated values vary with software revisions.
9.9 Sensor data ranges
Type of sensor Rated (Center) No readjustment necessary range
Max. automatic correction range
Pat. pressure 0.00 mbar
- 0.8 to + 0.8 mbar
- 3.0 to + 3.0 mbar
Exp. flow
500 digits
375 to 625 digits
25 to 975 digits
AIR flow
250 digits
175 to 325 digits
25 to 475 digits
O2 flow
250 digits
175 to 325 digits
25 to 475 digits
Содержание Centiva/5 Plus
Страница 1: ...GE Healthcare Centiva 5 Plus Technical Reference Manual ...
Страница 14: ...1 Introduction TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 1 8 Notes ...
Страница 20: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 6 ...
Страница 27: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 13 ...
Страница 30: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 16 Notes ...
Страница 34: ...3 Post service check out TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 3 4 Notes ...
Страница 70: ...4 Test and calibration TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 4 36 Notes ...
Страница 154: ...7 Repair procedures TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a 7 48 Notes ...
Страница 178: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 2 9 1 Pneumatic diagram ...
Страница 180: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 4 9 2 Electrical diagram ...
Страница 181: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 5 ...
Страница 182: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 6 9 3 Component layout PC board B ...
Страница 183: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 7 9 4 Component layout PC board A ...
Страница 192: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 16 Notes ...
Страница 206: ...10 Configuration and optionalization TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 10 14 Notes ...
Страница 212: ...11 Optional accessories TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 11 6 Notes ...
Страница 213: ...TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 Notes ...