5
Troubleshooting
TRM, Centiva/5 Plus, M1047448/CM-922026, rev. 3.50.a of 11/05
5-25
5.3.3 Messages during
O2 sensor calibration
Code Message
Type Set condition
Action
Reset condition
n/a No
signal:
wrong
gas or sensor
defective.
Advi
ce
If sensor signal does not
exceed 40% at calibration
with 99% O2
Ensure proper O2 gas supply
Or
Replace O2 sensor
Sensor signal
exceeds 40%at
calibration with 99%
O2
n/a Signal
low:
sensor used-up
or sensor failure.
Advi
ce
If sensor signal does not
reach minimum data
sheet value at 99%
Or
21% O2
Ensure proper O2 gas supply
Or
Ensure proper AIR gas
supply
Or
Replace O2 sensor
If sensor signal is in
between minimum
and maximum data
sheet values at 21%
with AIR
And
99% with O2
n/a Signal
high:
sensor failure or
PCB failure.
Advi
ce
If sensor signal exceeds
maximum data sheet
value at 99%
Or
21% O2
Ensure proper O2 gas supply
Or
Ensure proper AIR supply
Or
Replace O2 sensor
Or B board
If sensor signal is in
between minimum
and maximum data
sheet values at 21%
with AIR
And
99% with O2
n/a Calibration
successfully
finished.
Advi
ce
If calibration settles
properly at
21% with AIR
and
99% with O2
Finish the calibration
procedure
n/a
Содержание Centiva/5 Plus
Страница 1: ...GE Healthcare Centiva 5 Plus Technical Reference Manual ...
Страница 14: ...1 Introduction TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 1 8 Notes ...
Страница 20: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 6 ...
Страница 27: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 13 ...
Страница 30: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 16 Notes ...
Страница 34: ...3 Post service check out TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 3 4 Notes ...
Страница 70: ...4 Test and calibration TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 4 36 Notes ...
Страница 154: ...7 Repair procedures TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a 7 48 Notes ...
Страница 178: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 2 9 1 Pneumatic diagram ...
Страница 180: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 4 9 2 Electrical diagram ...
Страница 181: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 5 ...
Страница 182: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 6 9 3 Component layout PC board B ...
Страница 183: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 7 9 4 Component layout PC board A ...
Страница 192: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 16 Notes ...
Страница 206: ...10 Configuration and optionalization TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 10 14 Notes ...
Страница 212: ...11 Optional accessories TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 11 6 Notes ...
Страница 213: ...TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 Notes ...